SUNMED
Received Feb 21, 2022 · Event occurred Dec 15, 2021
Report 1314417-2022-00007 · MDR key 13573204
Device
Generic name
Small Adult Airflow W/manometer
Manufacturer
Sunmed Holdings LlcModel number
AF5140MBLot number
312679
Product problems
- Difficult to Open or Close
Patient
Unknown
- Insufficient Information
- Insufficient Information
Narrative
Additional Manufacturer Narrative
THE PATIENT COULD NOT RECEIVE PROPER RESPIRATORY SUPPORT AND WAS REPORTED TO BE UNRESPONSIVE. SUMMARY: THE CUTTING PROCESS USED FOR THE PATIENT VALVE DUCKBILL IS THE LIKELY ROOT CAUSE FOR THE DUCKBILL BEING SEALED. ON 5/10/2019, THE DRAWING FOR THE DUCKBILL WAS UPDATED TO ADD A BLADE SPECIFICATION TO REDUCE THE MATERIAL'S ABILITY TO RE-ADHERE AFTER CUTTING. RA: SEALED DUCKBILL HAS WORST-POTENTIAL SEVERITY OF 9 - HAZARDOUS. THIS COMPLAINT WILL BE BROUGHT TO CARB FOR REVIEW. HOWEVER, RISK IS ALREADY MITIGATED SIGNIFICANTLY WITH IN-PROCESS INSPECTION REQUIREMENTS, BLADE SPECIFICATION ADDED 5/10/2019, AND IFU REQUIREMENTS TO TEST THE RESUSCITATION BAG PRIOR TO USE.
Description of Event or Problem
DUCK BILL SEALED SHUT. CUSTOMER WAS UNRESPONSIVE, USED THE AF5140MB THE DUCK BILL VALVE WOULD NOT ALLOW THE AIR TO BE SQUEEZED OUT OF IT. PATIENT WOKE UP AFTERWARDS. THEY TRIED AGAIN AND WITH A LOT OF PRESSURE IT WORKED.
Additional Manufacturer Narrative
THE PATIENT COULD NOT RECEIVE PROPER RESPIRATORY SUPPORT AND WAS REPORTED TO BE UNRESPONSIVE. SUMMARY: THE CUTTING PROCESS USED FOR THE PATIENT VALVE DUCKBILL IS THE LIKELY ROOT CAUSE FOR THE DUCKBILL BEING SEALED. ON (B)(6) 2019, THE DRAWING FOR THE DUCKBILL WAS UPDATED TO ADD A BLADE SPECIFICATION TO REDUCE THE MATERIAL'S ABILITY TO RE-ADHERE AFTER CUTTING. RA: SEALED DUCKBILL HAS WORST-POTENTIAL SEVERITY OF 9 - HAZARDOUS. THIS COMPLAINT WILL BE BROUGHT TO CARB FOR REVIEW. HOWEVER, RISK IS ALREADY MITIGATED SIGNIFICANTLY WITH IN-PROCESS INSPECTION REQUIREMENTS, BLADE SPECIFICATION ADDED (B)(6) 2019, AND IFU REQUIREMENTS TO TEST THE RESUSCITATION BAG PRIOR TO USE.
Description of Event or Problem
DUCK BILL SEALED SHUT. CUSTOMER WAS UNRESPONSIVE, USED THE AF5140MB THE DUCK BILL VALVE WOULD NOT ALLOW THE AIR TO BE SQUEEZED OUT OF IT. PATIENT WOKE UP AFTERWARDS. THEY TRIED AGAIN AND WITH A LOT OF PRESSURE IT WORKED.