inforMED
MalfunctionFII

AQARIUS - AUTOMATED FLUID

Received Jan 23, 2009 · Event occurred Nov 26, 2008

Report 6000002-2009-09731 · MDR key 1352966

Device

Generic name

Acute Blood Purification Hemofiltration Machine

Model number

GEF08200

Catalog number

NA

Lot number

NA

Product problems

  • Inaccurate Flow Rate

Patient

UNK

  • No Consequences Or Impact To Patient

Narrative

Description of Event or Problem

REPORTEDLY, TAKING TOO MUCH FLUID FROM PATIENT.

Additional Manufacturer Narrative

EVALUATION SUMMARY: RECALIBRATED ALL PRESSURES AND SCALES. MACHINE PASSING SELF TEST. SUBSTITUTION SCALE. SYSTEM. FILTRATE SCALE. SYSTEM. FILTRATE PRESSURE SENSOR. SYSTEM. ACCESS PRESSURE SENSOR. SYSTEM - CALIBRATION ERROR.