inforMED
MalfunctionFPB

MEDEX MICRO BORE EXTENSION SETS

Received Feb 14, 2022 · Event occurred Apr 28, 2021

Report 3012307300-2022-03351 · MDR key 13529352

Device

Generic name

Filter, Infusion Line

Model number

FS116

Catalog number

FS116

Lot number

4092417

Product problems

  • Gas/Air Leak
  • Gas/Air Leak

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE DEVICE WAS RETURNED TO SMITHS MEDICAL. A SAMPLE WAS TEST FOR LEAKAGE BY MANUALLY INTRODUCING WATER USING A SYRINGE. THE SAMPLE PRESENT LEAKAGE BY ONE OF THE SIDE OF THE FILTER. IT WAS OBSERVED THAT THE FILTER IS OPEN IN ONE SIDE WHERE SHOULD BE WELDED. IT PRESENT DAMAGED WHICH MAY INDICATE A WRONG HANDLING OF THE PART. DHR REVIEW WAS EXECUTED AND NO DISCREPANCIES WERE OBSERVED DURING THE MANUFACTURING OF REPORTS LOT 4068818. THE LOT USED OF FILTER P1123-07 WAS 6447/20, NO DEFECTS WERE REPORTED ON THIS LOT OF FILTER. COMPLAINT IS CONFIRMED DUE TO SAMPLE RETURN PRESENT LEAKAGE AT FILTER, HOWEVER ROOT CAUSE IS NOT ATTRIBUTABLE TO MANUFACTURING.

Description of Event or Problem

IT WAS REPORTED THAT THE PICU HAS HAD "7-8+" OCCURRENCES OF CRACKED/LEAKING FILTERS FROM THE SAME LOT. NO PATIENT INJURY WAS REPORTED.