inforMED
MalfunctionDWH

UNIVERSAL HOSE, 10'

Received Mar 13, 2009

Report 1017294-2009-00039 · MDR key 1347151

Device

Generic name

Surgical Instrument Motor Access/attachment

Manufacturer

Conmed Linvatec

Model number

NA

Catalog number

00505201000

Lot number

NK

Product problems

  • Explode

Patient

Not reported

  • Treatment with medication(s)

Narrative

Additional Manufacturer Narrative

INVESTIGATION RESULTS: AN EVALUATION OF THE RETURNED HOSE FOUND THE OUTER HOSE HAD RUPTURED AT THE COUPLING END. FURTHER EXAMINATION FOUND THE DIFFUSER END DAMAGED (OUT OF ROUND) AND EVIDENCE OF THIRD PARTY REPAIR. USER MEDWATCH RECEIVED FROM FDA ON FEBRUARY 23, 2009. THIS EVENT IS BEING REPORTED BECAUSE THE HOSE WAS ASSOCIATED WITH A FAILURE WHICH RESULTED IN MEDICAL INTERVENTION (ANTIBIOTICS GIVEN).