ERGOSTYLE FX WITH MANUAL DROPS

Received Jan 30, 2009

Report 1022819-2009-00021 · MDR key 1343363

Device

Manufacturer

Chattanooga Group

Model number

5820

Catalog number

5820

Product problems

  • Break
  • Burst Container or Vessel
  • Failure to Shut Off

Patient

Not reported

  • No Information

Narrative

Description of Event or Problem

THE CLINICIAN WAS LOWERING THE TABLE SURFACE WITH THE PATIENT ON THE TABLE SURFACE WHEN THE MOTOR MALFUNCTIONED. THE TABLE REACHED FULL RETRACTION AND THE MOTOR CONTINUED TO ATTEMPTING TO LOWER THE TABLE SURFACE. THIS ACTION CAUSED THE UPPER YOKE WHERE THE MOTOR IS SECURED TO FLEX. THE FLEX MOTION BROKE THE CASTING OF THE ACTUATOR. THE BROKEN PIECE OF CASTING DID EJECT FROM UNDER THE TABLE.

Additional Manufacturer Narrative

AWAITING DEVICE RETURN AND EVALUATION. ONCE THE DEVICE IS RETURNED AND EVALUATED, A CONCLUDING FORM 3500A WILL BE SENT TO THE FDA-MDR.