LUCEA 40
Received Feb 2, 2022 · Event occurred Jan 7, 2022
Report 9710055-2022-00054 · MDR key 13423298
Device
Generic name
Device, Medical Examination, Ac Powered
Manufacturer
Maquet SasModel number
ARD568601997Catalog number
ARD568601997
Product problems
- Crack
- Crack
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
ADDITIONAL INFORMATION WILL BE PROVIDED FOLLOWING THE CONCLUSION OF THE INVESTIGATION. DEVICE NOT RETURNED TO MANUFACTURER.
Description of Event or Problem
ON (B)(6) 2022 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF EXAMINATION LIGHTS - LUCEA 40. AS IT WAS STATED, THE LAMP HAVE BROKEN HOUSING. PHOTO EVIDENCE SHOWED, THAT SOME PARTICLES WERE MISSING. THERE WAS NO INJURY REPORTED HOWEVER WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTS FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MIGHT CAUSE A CONTAMINATION.
Description of Event or Problem
MANUFACTURER REFERENCE NUMBER (B)(4).
Additional Manufacturer Narrative
GETINGE BECAME AWARE OF AN ISSUE WITH EXAMINATION LIGHT - LUCEA 40. AS IT WAS STATED, THE LAMP HAS BROKEN HOUSING. PHOTO EVIDENCE SHOWED THAT SOME PARTICLES WERE MISSING. THERE WAS NO INJURY REPORTED, HOWEVER WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTS FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MIGHT CAUSE A CONTAMINATION. BASED ON SSU TECHNICIAN STATEMENT, SOME OF THE AFFECTED DEVICES WERE MODIFIED AND BROUGHT BACK TO SERVICE. IT WAS ESTABLISHED THAT WHEN THE EVENT OCCURRED, THE SURGICAL LIGHT DID NOT MEET ITS SPECIFICATION DUE TO CRACKS AND MISSING PARTICLES FROM COVERS, WHICH CONTRIBUTED TO THE EVENT. PROVIDED INFORMATION DO NOT INDICATE IF UPON THE EVENT OCCURRENCE THE DEVICE WAS BEING USED FOR PATIENT TREATMENT. REVIEW OF RECEIVED CUSTOMER PRODUCT COMPLAINTS RELATED TO INVESTIGATED ISSUE, REVEALED THAT THERE WERE NO INJURIES TO A USER NOR TO A PATIENT OR OPERATOR WHEN THIS PARTICULAR MALFUNCTION OCCURRED. WE CAN ASSUME THAT THE FAILURE RATIO OF CRACKS AND MISSING PARTICLES ON LUCEA 10/40 DEVICES IS LOW. AS STATED BY SUBJECT MATTER EXPERT AT MANUFACTURER¿S, CRACKS WERE DETECTED AT THE EDGE OF FIXING HOLES OF TRANSPARENT HOUSING AND HANDLE INTERFACE WERE PROBABLY CAUSED BY THE INCOMPATIBILITY OF THE CLEANING PROTOCOL OR AN ABNORMAL USE. THE USER MANUAL FOR LUCEA 10/40 DESCRIBES HOW TO CLEAN AND DISINFECT THE LIGHT HEADS. THIS DOCUMENT INCLUDES SOME OF THE RECOMMENDED AND PROHIBITED PRODUCTS. IN ORDER TO AVOID MECHANICAL STRESSES APPLIED ON THE TRANSPARENT HOUSING DURING USE THE USER MANUAL MENTIONS TO HANDLE THE LIGHT BY THE HANDLE. TO PREVENT SIMILAR EVENT, THE SAME USER MANUAL MENTIONS TO CHECK THE LIGHT HEADS DURING DAILY INSPECTION. WE BELIEVE THE RELATED DEVICES ARE PERFORMING CORRECTLY IN THE MARKET. WE ALSO BELIEVE THAT IF THE MANUFACTURER RECOMMENDATION WOULD HAVE BEEN FOLLOWED THE INCIDENT WOULD HAVE BEEN AVOIDED. GETINGE SHALL CONTINUE TO MONITOR FOR ANY FURTHER EVENTS OF THIS NATURE AND DO NOT PROPOSE ANY FURTHER ACTION AT THIS TIME.
Additional Manufacturer Narrative
A FOLLOW-UP REPORT IS REQUIRED DUE TO NEW INFORMATION RECEIVED. THE CORRECTION OF D4 CATALOG # AND VERSION OF MODEL # DEEMS REQUIRED. THIS IS BASED ON THE INTERNAL EVALUATION. PREVIOUS CATALOG #:ARDLCA209021C; CORRECTED CATALOG #:ARD568601997; PREVIOUS VERSION OR MODEL #:ARDLCA209021C; CORRECTED VERSION OR MODEL #:ARD568601997. FURTHER INFORMATION PROVIDED BY GETINGE EMPLOYEE INDICATED THAT CRACKS WITHOUT MISSING PARTICLES OCCURRED ON DEVICE. BASED ON ADDITIONAL INPUT FROM GETINGE EMPLOYEE IT WAS POSSIBLE TO DETERMINE THAT THE ISSUE INVESTIGATED HEREIN IS NOT SAFETY AND RISK RELATED, AS THERE WAS NO INDICATION OF MISSING PARTICLES, WHICH WAS INITIALLY CONSIDERED. THEREFORE, THE SCENARIO DESCRIBED IN THE RECORD IS CONSIDERED AS NON-REPORTABLE. IT IS NOT LIKELY THAT THE ISSUE COULD LEAD TO THE SERIOUS INJURY OR DEATH WHEN THE MALFUNCTION REOCCURS. THAT IS WHY IT IS NOT CONSIDERED TO BE SAFETY RELATED AND REPORTABLE.
Description of Event or Problem
MANUFACTURER REFERENCE NUMBER (B)(4).