AXSYM RUBELLA LGG ANTIBODY

Received Mar 12, 2009 · Event occurred Feb 7, 2008

Report 1415939-2008-00294 · MDR key 1341440

Device

Generic name

Meia For The Measurement Of Lgg Antibody To The Rubella Virus

Model number

NA

Catalog number

3B23-20

Lot number

56323M101

Product problems

  • Calibration Error
  • Device Displays Incorrect Message

Patient

Not reported

  • No Consequences Or Impact To Patient

Narrative

Additional Manufacturer Narrative

THIS IS AN INITIAL REPORT. AN INVESTIGATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.

Description of Event or Problem

THE CUSTOMER CONTACTED ABBOTT REPORTING CALIBRATION FAILURE FOR THE AXSYM RUBELLA LGG ASSAY, WHERE ERROR CODE 1111 (CALIBRATION CHECK FAILURE, M-CAL 1, RESPONSE TOO LARGE) WAS GENERATED. THE CUSTOMER STATED THE RUBELLA REAGENT AND CALIBRATORS WERE NEW. THE CUSTOMER REVIEWED TROUBLESHOOTING PROCEDURES WITH ABBOT REPRESENTATIVE AND WAS ADVISED TO CENTRIFUGE THE CALIBRATORS. THE CUSTOMER REPORTED RECALIBRATION WAS SUCCESSFUL AND QUALITY CONTROLS WERE WITHIN RANGE. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED BY THE CUSTOMER.

Remedial action

  • Other