AXSYM RUBELLA LGG ANTIBODY
Received Mar 12, 2009 · Event occurred Jan 30, 2008
Report 1415939-2008-00288 · MDR key 1341434
Device
Generic name
Meia For The Measurement Of Lgg Antibody To The Rubella Virus
Manufacturer
Abbott LaboratoriesModel number
NACatalog number
3B23-20
Lot number
56323M101
Product problems
- Calibration Error
- Device Displays Incorrect Message
Patient
Not reported
- No Consequences Or Impact To Patient
Narrative
Description of Event or Problem
THE CUSTOMER CONTACED ABBOTT REGARDING THREE ATTEMPTS TO CALIBRATE THE AXSYM RUBELLA LGG ASSAY WHICH FAILED, AND ERROR CODE 1111 (CALIBRATION CHECK FAILURE, M-CAL 1, RESPONSE TOO LARGE) WAS GENERATED. THE CUSTOMER REPORTED AXSYM MAINTENANCE WAS UP TO DATE. FOR TROUBESHOOTING, THE CUSTOMER ATTEMPTED CALIBRATION USING WATER INSTEAD OF MASTER CALIBRATOR 1, WHICH WAS SUCCESSFUL. THE CUSTOMER CLEANED THE INSTRUMENT OPTICS AND ATTEMPTED RECALIBRATION WHICH FAILED AGAIN WITH THE SAME ERROR MESSAGE. THE CUSTOMER WAS SENT NEW REAGENTS AND CALIBRATORS AND UPON RECALIBRATION WITH THE NEW MATERIAL, THE SAME CALLIBRATION FAILURE PERSISTED. THE CUSTOMER WAS ADVISED TO CENTRIFUGE CALIBRATOR A, THEN USE THE CENTRIFUGED CALIBRATOR IN THE RUBELLA RECALIBRATION WHICH WAS SUCCESSFUL. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED BY THE CUSTOMER.
Additional Manufacturer Narrative
THIS IS AN INITIAL REPORT. AN INVESTIGATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.
Remedial action
- Other