inforMED
MalfunctionDRG

VITAL SYNC REMOTE MONITORING SYSTEM

Received Jan 24, 2022 · Event occurred Jan 7, 2022

Report 1717344-2022-00079 · MDR key 13333156

Device

Generic name

Transmitters And Receivers, Physiological Sign

Model number

10116080

Catalog number

10116080

Product problems

  • Computer Software Problem
  • Computer Software Problem

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem

ACCORDING TO THE REPORTER, PRIOR TO USE, THE REMOTE SURVEILLANCE SYSTEM WAS NOT ALARMING. MULTIPLE DEVICES ON DIFFERENT FLOORS WERE NOT ALARMING. THERE WERE NO ERRORS WITH THE REMOTE SURVEILLANCE SYSTEM, AND THE DATA COLLECTION SERVER WAS RESTARTED. IT WAS FOUND THAT THE ISSUE WAS TIED UP TO THE HEALTHCARE DATA INTEGRATION ENGINE. IT GENERATED AN ADDITIONAL LINE ITEM AND CAUSED THE ALARMS NOT TO PROCESS. A LINE ITEM WAS ADDED TO HEALTHCARE DATA INTEGRATION ENGINE WHICH REMOVED THE ADDITIONAL LINE THAT CAME FROM THE THIRD PARTY SYSTEM. THERE WAS NO PATIENT INVOLVEMENT.