HOT DOG PATIENT WARMING SYSTEM MULTIFUNCTION CONTROLLER
Received Dec 21, 2021 · Event occurred Nov 16, 2021
Report 13051708 · MDR key 13051708
Device
Generic name
System, Thermal Regulating
Manufacturer
Augustine Temperature Management LlcModel number
WC52Catalog number
WC52
Product problems
- Insufficient Device Problem Information
Patient
30295 DA · Male
- Superficial (First Degree) Burn
- Superficial (First Degree) Burn
Narrative
Description of Event or Problem
DURING THIS SURGERY, CLINICAL STAFF RECEIVED A CALL FROM THE POST-ANESTHESIA CARE UNIT ABOUT A "WELT" ON THE PATIENTS LEFT KNEE. DOCTOR WAS GOING TO CHECK ON THE PREVIOUS PATIENTS KNEE AFTER THIS SURGERY WAS COMPLETED. DOCTOR DETERMINED IT WAS A BURN ON THE LEFT KNEE ON THE PREVIOUS PATIENT. CLINICAL STAFF CHECKED THE HOTDOG PATIENT WARMING SYSTEM ON THE CURRENT PATIENT AND IT WAS VERY HOT TO THE TOUCH. CLINICAL STAFF TOOK IT OFF AND SAW THE PATIENT'S KNEES LOOKED TO BE BURNED AS WELL. THE READING ON THE HOTDOG DEVICE WAS 41 DEGREES WHICH IS NORMAL. THE MACHINE AND THE BLANKET WERE BOTH GOOD ON DATES AND WERE NOT EXPIRED. A BLANKET WAS PLACED ON BOTH PATIENTS CREATING A BARRIER BETWEEN THE HOTDOG BLANKET AND PATIENTS SKIN. THE MACHINE AND BLANKET WERE BOTH REMOVED FROM THE ROOM AND SENT DOWN TO BE LOOKED AT.