inforMED
InjuryOUR

IFUSE IMPLANT SYSTEM

Received Nov 29, 2021 · Event occurred Nov 1, 2021

Report 3007700286-2021-00124 · MDR key 12887921

Device

Generic name

Orthopedic Rod

Manufacturer

Si-bone, Inc.

Model number

7035M-90

Lot number

2659341

Product problems

  • Malposition of Device
  • Malposition of Device

Patient

76 YR · Male

  • Undesired Nerve Stimulation
  • Undesired Nerve Stimulation

Narrative

Additional Manufacturer Narrative

BASED ON THE INFORMATION PROVIDED, REVIEW OF THE SURGICAL TECHNIQUE MANUAL, IFU, CERTIFICATES OF CONFORMANCE AND FMEA, THERE IS NO INDICATION OF DEVICE FAILURE AND NO INDICATION THAT THE DEVICE WAS OUT OF SPECIFICATION. THE MOST PROBABLE ROOT CAUSE IS USER ERROR: USING TOO LONG OF AN IMPLANT OR INSTALLING THE IMPLANT TOO DEEP. ADDITIONALLY, PER THE SURGICAL TECHNIQUE MANUAL (C-ARM AND O-ARM), THE SURGEON IS INSTRUCTED AS FOLLOWS: "PRIOR TO CLOSURE, ALWAYS OBTAIN FINAL FLUOROSCOPIC IMAGES IN THE LATERAL, INLET, AND OUTLET VIEWS TO CONFIRM NO CORTICAL WALL BREACH, FORAMEN BREACH, OR OTHER MALPOSITION."

Description of Event or Problem

THE PATIENT HAD LEFT SIDE SI JOINT ARTHRODESIS IN (B)(6) 2021 WHERE THREE IMPLANTS WERE INSTALLED. THE PATIENT COMPLAINED OF BUTTOCK PAIN POST OP AND PAIN RADIATING DOWN LEG. THE INFERIOR IMPLANT WAS IMPINGING ON THE NEUROFORAMEN. IN (B)(6) 2021, THE SURGEON PERFORMED A REVISION WHERE HE REMOVED THE INFERIOR-POSITIONED IMPLANT. NO BONE GRAFT OR ADDITIONAL HARDWARE WAS ADDED. NONE OF THE OTHER PREEXISTING IMPLANTS WERE ADJUSTED OR REMOVED. THE STATUS OF THE PATIENT FOLLOWING THE REVISION PROCEDURE IS NOT KNOWN.