inforMED
InjuryOUR

IFUSE IMPLANT SYSTEM

Received Nov 23, 2021 · Event occurred Oct 26, 2021

Report 3007700286-2021-00118 · MDR key 12862255

Device

Generic name

Orthopedic Rod

Manufacturer

Si-bone, Inc.

Model number

7065M

Lot number

493428

Product problems

  • Malposition of Device
  • Malposition of Device

Patient

52 YR · Male

  • Undesired Nerve Stimulation
  • Undesired Nerve Stimulation

Narrative

Description of Event or Problem

THE PATIENT HAD RIGHT SIDE SI JOINT ARTHRODESIS IN (B)(6) 2019 WHERE THREE IMPLANTS WERE INSTALLED. LATER THE PATIENT COMPLAINED OF RADICULAR PAIN. THE SURGEON DETERMINED THAT THE SUPERIOR-POSITIONED IMPLANT MAY HAVE BEEN IMPINGING ON THE NEUROFORAMEN. IN (B)(6) 2021, THE SURGEON PERFORMED A REVISION PROCEDURE WHERE HE REMOVED THE SUPERIOR POSITIONED IMPLANT USING CHISLES. THE IMPLANT WAS REPLACED WITH A SHORTER IMPLANT OF THE SAME TYPE IN A MORE POSTERIOR POSITION. NO OTHER PREEXISTING IMPLANTS WERE ADJUSTED OR REMOVED. THE STATUS OF THE PATIENT FOLLOWING THE REVISION PROCEDURE IS NOT KNOWN.

Additional Manufacturer Narrative

BASED ON THE INFORMATION PROVIDED, REVIEW OF THE SURGICAL TECHNIQUE MANUAL, IFU, CERTIFICATES OF CONFORMANCE AND FMEA, THERE IS NO INDICATION OF DEVICE FAILURE AND NO INDICATION THAT THE DEVICE WAS OUT OF SPECIFICATION. THE MOST PROBABLE ROOT CAUSE IS USER ERROR: USING TOO LONG OF AN IMPLANT OR INSTALLING THE IMPLANT TOO DEEP. ADDITIONALLY, PER THE SURGICAL TECHNIQUE MANUAL, THE SURGEON IS INSTRUCTED AS FOLLOWS: "PRIOR TO CLOSURE, ALWAYS OBTAIN FINAL FLUOROSCOPIC IMAGES IN THE LATERAL, INLET, AND OUTLET VIEWS TO CONFIRM NO CORTICAL WALL BREACH, FORAMEN BREACH, OR OTHER MALPOSITION."