CORPATH GRX
Received Nov 23, 2021 · Event occurred Oct 29, 2021
Report 3007822508-2021-00005 · MDR key 12860977
Device
Generic name
Corpath Grx System Power Cord
Manufacturer
Corindus IncProduct problems
- Power Problem
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THERE WAS NO PATIENT OR USER INJURY REPORTED. BASED ON THE DESCRIPTION OF EVENTS THE SUBSTANCE THAT LEAKED FROM THE IV BAG WAS ELECTROLYTIC AND WAS REPEATEDLY CONTACTING THE POWER INPUT. WHEN AN ELECTROLYTIC SOLUTION IS PRESENT ON AN AREA OF THE POWER INPUT A REACTION CAN OCCUR THAT RESULTS IN THE ISSUE THAT WAS REPORTED. THE POWER SUPPLY CONTAINED THE REACTION AS DESIGNED BY PREVENTING DAMAGE TO THE ROBOT OR THE USER OR PATIENT EXPERIENCING HARM.
Description of Event or Problem
THE POWER SUPPLY NO LONGER FUNCTIONED AFTER AN IV BAG LEAKED FLUID ONTO IT.