inforMED
MalfunctionPRI

VIDAS BRAHMS PROCALCITONIN 60T

Received Nov 19, 2021 · Event occurred Oct 21, 2021

Report 8020790-2021-00278 · MDR key 12847815

Device

Generic name

Vidas Brahms Procalcitonin 60t

Manufacturer

Biomerieux Sa

Catalog number

30450-86

Lot number

1008741810

Product problems

  • Unable to Obtain Readings
  • Unable to Obtain Readings

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

FOLLOWING AN INCREASE IN CUSTOMER COMPLAINTS ABOUT ¿SUBSTRATE ERROR¿ ON DIFFERENT VIDAS® REFERENCES TESTED ON ALL VIDAS SYSTEMS, AN INTERNAL INVESTIGATION WAS PERFORMED. INVESTIGATION: **DEVICE HISTORY RECORD** THE ANALYSIS OF THE BATCH HISTORY RECORDS OF THE VIDAS LOTS CONCERNED SHOWED NO ANOMALY DURING THE MANUFACTURING, CONTROL, AND PACKAGING PROCESSES. NO NON-CONFORMITY LINKED TO THE CUSTOMER¿S COMPLAINT WAS REGISTERED ON THESE BATCHES. FOLLOWING THE INCREASE OF COMPLAINTS FOR SUBSTRATE ISSUES, A CAPA (CORRECTIVE ACTION PREVENTIVE ACTION) HAS BEEN OPENED TO IDENTIFY THE ACTIONS. **TESTS/ANALYSIS PERFORMED** - THE PROBLEM WAS DETECTED ON VIDAS, WHEN THE FLUORESCENCE IS MEASURED (FIRST READ) BEFORE THE BEGINNING OF THE TESTS. IF THERE IS A ¿SUBSTRATE ERROR¿, THE USER HAS AN ERROR MESSAGE ¿ ERRORS/SUBSTRATE ERROR¿ WHICH DISPLAYS ON THE EQUIPMENT MAKING IT IMPOSSIBLE TO RUN TESTS AND GENERATE RESULTS. - THE ¿SUBSTRATE ERROR¿ ISSUE HAS BEEN OBSERVED 5 TO 6 MONTHS AFTER THE DATE OF MANUFACTURING OF THE STRIPS. - DURING THE INVESTIGATION, IT WAS CONFIRMED THAT ONE LOT OF COMMON RAW MATERIAL WAS USED IN ALL SUBSTRATE BATCHES CONTAINED IN VIDAS® IMMUNO-ASSAYS STRIPS WITH A HIGH LEVEL OF COMPLAINTS ABOUT ¿SUBSTRATE ERROR¿ AND THEREFORE THIS BATCH OF RAW MATERIAL WAS IDENTIFIED AS THE MOST PROBABLE COMMON ROOT-CAUSE. - INVESTIGATION PERFORMED AT THE SUPPLIER LEVEL CONFIRMED THAT A LOT OF RAW MATERIAL IN CAUSE IS ATYPICAL AND IS INDEED THE SOURCE OF SUBSTRATE ERRORS ON VIDAS REAGENTS. - AT THE END OF SEPTEMBER 2021, THE MANUFACTURING SITE STOPPED THE PRODUCTION OF THE KITS WITH THE CONCERNED RAW MATERIAL LOT. **ROOT CAUSE ANALYSIS AND CONCLUSION** - THE ROOT CAUSES LINKED TO THE SPECIFICATIONS PROCESS AT THE SUPPLIER OF THE RAW MATERIAL HAVE BEEN IDENTIFIED. - CORRECTIVE ACTION WILL BE IMPLEMENTED AT THE SUPPLIER TO ADD SOME SPECIFICATIONS AND CONTROL IN THE MANUFACTURING PROCESS OF THE RAW MATERIAL CONCERNED. - AT BIOMÉRIEUX SITE, AN INCOMING CONTROL WITH STABILITY WILL BE IMPLEMENTED AS A PREVENTIVE ACTION.

Description of Event or Problem

INTENDED USE: VIDAS® B·R·A·H·M·S PCT¿ IS AN AUTOMATED TEST FOR USE ON THE VIDAS FAMILY INSTRUMENTS FOR THE DETERMINATION OF HUMAN PROCALCITONIN IN HUMAN SERUM OR PLASMA (LITHIUM HEPARIN) USING THE ELFA (ENZYME-LINKED FLUORESCENT ASSAY) TECHNIQUE. THE VIDAS® B·R·A·H·M·S PCT¿ IS INTENDED FOR USE IN CONJUNCTION WITH OTHER LABORATORY FINDINGS AND CLINICAL ASSESSMENTS TO AID IN THE RISK ASSESSMENT OF CRITICALLY ILL PATIENTS ON THEIR FIRST DAY OF ICU ADMISSION FOR PROGRESSION TO SEVERE SEPSIS AND SEPTIC SHOCK. DESCRIPTION OF THE ISSUE: ON (B)(6) 2021, A CUSTOMER FROM RUSSIA NOTIFIED BIOMÉRIEUX OF OBTAINING SUBSTRATE ERROR LEADING TO DELAYED RESULT WHILE USING VIDAS® BRAHMS PROCALCITONIN 60T (REF. 30450-86, LOT #1008741810) WITH PATIENT SAMPLES. AN ERROR MESSAGE IS DISPLAYED ON THE EQUIPMENT VISIBLE BY THE USER, AND IT IS IMPOSSIBLE TO RUN TESTS AND GENERATE RESULTS. THE CUSTOMER NOTIFIED BIOMÉRIEUX OF A DELAY, HOWEVER, AT THIS TIME OF THE COMPLAINT, THE LENGTH OF THE DELAY AND THE NUMBER OF IMPACTED PATIENT IS UNKNOWN. THERE IS NO INDICATION OR REPORT FROM THE LABORATORY THAT THE ISSUE LED TO ANY ADVERSE EVENT RELATED TO ANY PATIENT'S STATE OF HEALTH. IMPACTED LOT IS PART OF THE FSCA5333 RELEASED ON (22-SEP-2021 FOR SUBSTRATE ERROR ON MULTIPLE VIDAS PRODUCT REFERENCE. A BIOMÉRIEUX INTERNAL INVESTIGATION HAS BEEN INITIATED. NOTE: REFERENCE 30450-86 IS NOT REGISTERED IN THE UNITED STATES. THE U.S. SIMILAR DEVICE IS PRODUCT REFERENCE 30450-01.

Additional Manufacturer Narrative

SINCE JULY 2021, BIOMÉRIEUX HAS BEEN RECEIVING AN INCREASING NUMBER OF COMPLAINTS LINKED TO A VIDAS® ¿SUBSTRATE ERROR¿. IT PREVENTS THE TEST FROM BEING RUN, THEREFORE LEADS TO POTENTIAL DELAYED RESULTS AS THE USER MUST RUN ANOTHER TEST. A MEASUREMENT OF THE BACKGROUND NOISE SIGNAL (RFU) IS MADE BY THE VIDAS® SYSTEM PRIOR TO LAUNCHING THE REACTION. AN ACCEPTABLE LIMIT IS DEFINED DURING PRODUCT DESIGN FOR EACH REFERENCE OF FINISHED GOODS. THE ACCEPTABLE LIMIT IS 350 RFU FOR BRAHMS PROCACLITONIN AND 500 RFU FOR THE OTHER US IMPACTED REFERENCES. THE SUBSTRATE IS PRESENT IN THE LAST WELL OF THE STRIP OF ALL VIDAS® IMMUNO-ASSAYS AND ALLOWS FLUORESCENCE WHEN DEGRADED BY THE ENZYME (PAL). THE LEVEL OF FLUORESCENCE IS THEN CORRELATED WITH THE RESULTS OF ALL TESTS. WHEN PERFORMING A TEST, IF THE RFU IS HIGHER THAN THE ACCEPTABLE LIMIT DEFINED DURING THE PRODUCT DEVELOPMENT, THERE IS AN ERROR MESSAGE DISPLAYED BY THE SYSTEM: ¿SUBSTRATE ERROR¿. THE TEST IS STOPPED AND THIS ALARM PREVENTS THE SYSTEM FROM PROVIDING ANY RESULT. WITH THIS ALARM BEING PRESENT ON ALL THE SYSTEMS OF THE VIDAS® FAMILY, IT GUARANTEES THAT NO FALSE RESULTS CAN BE GIVEN IN CASE OF A SUBSTRATE DEGRADATION. THERE IS A POTENTIAL RISK OF DELAYED RESULTS. DESCRIPTION OF THE INVESTIGATION AND ROOT CAUSE ANALYSIS : INVESTIGATIONS WERE IMMEDIATELY INITIATED TO IDENTIFY THE ROOT-CAUSE, THE FOLLOWING WERE IDENTIFIED: ALL LOTS IMPACTED OF VIDAS® IMMUNO-ASSAYS WERE CONFORM TO THE SPECIFICATIONS AT RELEASE. THE SUBSTRATE ERROR ISSUE WAS CONFIRMED ON ALL LOTS OF VIDAS® IMMUNO-ASSAYS MANUFACTURED WITH SUBSTRATE BATCHES USING A COMMON LOT OF RAW MATERIAL (4-MUP) THAT WAS IDENTIFIED AS THE MOST PROBABLE COMMON ROOT-CAUSE. THE SCOPE OF THE ISSUE WAS IDENTIFIED ON ALL LOTS OF VIDAS® IMMUNO-ASSAYS MANUFACTURED SINCE FEBRUARY 2021 USING SUBSTRATE BATCHES MANUFACTURED WITH THIS CONCERNED LOT OF RAW MATERIAL. THE PROBLEM IS DUE TO AN ACCELERATED DEGRADATION OF THE SUBSTRATE. IT FOLLOWS A LINEAR MODEL OVER TIME LEADING TO RFU ACCEPTABLE LIMITS BEING REACHED BEFORE THE END OF THE REGISTERED SHELF LIFE OF THE PRODUCT. KINETIC EVOLUTION ANALYSES WERE PERFORMED BY MEASURING SUBSTRATE RFU OF A STATISTICALLY REPRESENTATIVE NUMBER OF VIDAS® IMMUNO- ASSAYS RETAINED BATCHES (MANUFACTURED WITH THE SUBSTRATE CONTAINING THE CONCERNED RAW MATERIAL) AT DIFFERENT SHELF-LIVES. THE MODEL WAS VALIDATED ON NUMEROUS DATA (~ 450 000) COLLECTED FROM THE FIELD (CUSTOMERS) . THE ANALYSIS OF THE KINETIC MODEL ALLOWS US TO PREDICT THE DEGRADATION TREND OF THE SUBSTRATE USING THE CONCERNED BATCH OF 4MUP AND THEREFORE TO REVISE THE EXPIRY DATES FOR EACH LOT OF IMPACTED VIDAS® IMMUNO- ASSAYS FINISHED PRODUCTS. WHEN USED UNTIL THE REVISED EXPIRY DATE, THE PRODUCT CONTINUES TO PERFORM PER ITS WHEN AN ERROR MESSAGE ¿SUBSTRATE ERROR¿ IS DISPLAYED BY THE VIDAS® SYSTEM, THE TEST IS STOPPED AND THIS ALARM PREVENTS THE SYSTEM FROM PROVIDING ANY RESULT. THE INABILITY TO PROVIDE A RESULT IN A TIMELY MANNER (POTENTIAL DELAYED RESULTS) MAY HAVE AN IMPACT ON PATIENT MANAGEMENT DEPENDING ON WHETHER OR NOT THERE ARE CLEAR MANAGEMENT OR TREATMENT STRATEGIES THAT RELY SOLELY OR MOSTLY ON THE TEST RESULT ; WHETHER OR NOT THE MANAGEMENT STRATEGY CAN BE DELAYED, AND FOR HOW LONG ; WHETHER OR NOT THERE ARE ASSAYS FROM A LOT NOT IMPACTED BY THE ISSUE THAT ARE AVAILABLE ; WHETHER OR NOT THE LABORATORY IS ABLE TO USE AN ALTERNATIVE METHOD. THEREFORE, A CORRECTIVE ACTION INVOLVING A REVISED EXPIRATION DATE FOR ALL IMPACTED LOTS OF CLINICAL VIDAS® IMMUNO-ASSAYS PRODUCTS IS REQUIRED TO ENSURE THAT THE SPECIFIED PRODUCTS WILL CONTINUE TO PERFORM PER LABELLED PERFORMANCE SPECIFICATIONS. THE INVESTIGATION IS ONGOING.

Remedial action

  • Recall