SUPPORT ARM 177
Received Nov 12, 2021
Report 8010042-2021-02695 · MDR key 12802174
Device
Generic name
Support, Arm
Manufacturer
Maquet Critical Care AbModel number
SUPPORT ARM 177Catalog number
6481720
Product problems
- Fracture
- Fracture
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
IT WAS REPORTED THAT THE SUPPORT ARM BROKE. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF #: (B)(4).
Additional Manufacturer Narrative
NO REPAIR WAS REQUESTED AND NO PARTS WERE RETURNED. THE SUPPORT ARM IS A CASTING AND IT MOST PROBABLY DEVELOPED A CRACK AT AN EARLIER OCCASION EITHER BY OVERLOADING OR IMPACT AND THIS LED TO THE REPORTED BREAKING. THE SUPPORT ARM HAS BEEN SUCCESSFULLY TESTED FOR MECHANICAL STRENGTH AND IS DESIGNED ACCORDING TO STANDARD. PREVIOUS INVESTIGATIONS OF SIMILAR FAULTS LED TO A REDESIGN AND CHANGE OF THE MANUFACTURING PROCESS IN ORDER TO OBTAIN A HIGHER MECHANICAL STRENGTH OF THE SUPPORT ARM. THE SUPPORT ARM ASSOCIATED WITH THIS COMPLAINT HAD BEEN MANUFACTURED BEFORE THIS CHANGE WAS IMPLEMENTED IN PRODUCTION.