inforMED
InjuryBTM

VITAL SIGNS¿ FLOW INFLATING TAIL-END VALVE TRANSPORT CIRCUIT, FLOW-INFLATING TRA

Received Nov 10, 2021 · Event occurred Sep 12, 2021

Report 8030673-2021-00210 · MDR key 12788935

Device

Generic name

Hyperinflation Bag System

Manufacturer

Vyaire Medical

Catalog number

5102RVW

Lot number

UNKNOWN

Product problems

  • Mechanical Problem
  • Mechanical Problem

Patient

Unknown

  • Pneumothorax
  • Pneumothorax

Narrative

Additional Manufacturer Narrative

THE SUSPECT DEVICE HAS NOT BEEN RETURNED FOR EVALUATION. NO PHOTOS AVAILABLE. THEREFORE, NO ROOT CAUSE HAS BEEN DETERMINED. VYAIRE MEDICAL WILL SUBMIT A SUPPLEMENTAL REPORT IN ACCORDANCE WITH 21 CFR SECTION 803.56 IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem

THE CUSTOMER REPORTED THAT THE VITAL SIGNS¿ FLOW INFLATING TAIL-END VALVE TRANSPORT CIRCUIT, FLOW-INFLATING TRA EXPERIENCED VALVE CLOSES TOO EASILY WHICH CAUSED HIGH PRESSURES TO ENTER LUNGS AND RESULTED IN PNEUMOTHORAX. AS AN INTERVENTION, INTUBATED INFANT BEING GIVEN MANUAL BREATHS.