MalfunctionECX
CRYO TANK
Received Nov 9, 2021 · Event occurred Oct 14, 2021
Report 12778516 · MDR key 12778516
Device
Generic name
Cryo Tank
Manufacturer
Praxair Distribution, Inc.Product problems
- Gas Output Problem
- Manufacturing, Packaging or Shipping Problem
- Therapeutic or Diagnostic Output Failure
Patient
Unknown
- Insufficient Information
- Insufficient Information
Narrative
Description of Event or Problem
AFTER CHANGING OUT THE CRYO TANK, THE CRYOABLATION MACHINE STOPPED WORKING. IT WAS FOUND THAT NITRIC OXIDE TANK CONTAINS THE WRONG GAS. THE CYRO TANK SHOULD HAVE NON-MEDICAL GRADE AIR. IT CONTAINED MEDICAL GRADE NITRIC OXIDE, WHICH IS USED BY ANESTHESIA. BOTH TANKS WERE APPROPRIATELY LABELED AND ARE SPECIFIC TO THE CRYOABLATION CONSOLE. THIS INCORRECT GAS CAUSED THE CONSOLE TO MALFUNCTION. ONCE THE CORRECT GAS FROM ANOTHER TANK WAS HOOKED UP AND TESTED, THE MACHINE WORKED FINE.