SALINE-FILLED TESTICULAR PROSTHESIS

Received Dec 31, 2008

Report 2125050-2008-00012 · MDR key 1276470

Device

Generic name

Testicular Prosthesis

Model number

5207501000

Catalog number

5207501000

Lot number

5712686-015

Product problems

  • Filling Problem
  • Device Damaged Prior to Use
  • Out-Of-Box Failure

Patient

Not reported

  • Surgical procedure, delayed

Narrative

Description of Event or Problem

AS REPORTED TO COLOPLAST, A PHYSICIAN REPORTED A PROSTHESIS WAS DAMAGED UPON RECEIPT, SO IT COULD NOT BE FILLED JUST PRIOR TO IMPLANT SURGERY, WHICH HAD TO BE POSTPONED.

Additional Manufacturer Narrative

ONE TESTICULAR DEVICE WAS RECEIVED FOR EVALUATION. EXAMINATION AND TESTING OF THE DEVICE REVEALED A SEPARATION ON THE SUTURE TAB. MICROSCOPIC EXAMINATION OF THE SEPARATION REVEALED A SMOOTH AND STRAIGHT EDGE INDICATING SHARP INSTRUMENTATION WAS INVOLVED. BECAUSE THESE COMPONENTS WERE RELEASED ACCORDING TO MANUFACTURING AND QUALITY CONTROL PROCEDURES, QA CONCLUDED THAT THE OBSERVED INSTRUMENT SEPARATION ON THE SUTURE TAB OCCURRED SUBSEQUENT TO THE DEVICE PACKAGING BEING OPENED. AS THE DEVICE WAS NEVER IMPLANTED. QA CONCLUDED THAT INSTRUMENT DAMAGE MOST LIKELY OCCURRED DURING THE DEVICE PREPPING PROCESS PRIOR TO IMPLANT.