13 GA TAPERED CANNULA, TROCAR STYLET, BEVEL STYLET
Received Nov 3, 2021 · Event occurred Oct 11, 2021
Report 9612164-2021-04192 · MDR key 12744710
Device
Generic name
Device, Percutaneous, Biopsy
Manufacturer
Medtronic MexicoModel number
VPB13DCatalog number
VPB13D
Lot number
221828512
Product problems
- Detachment of Device or Device Component
- Detachment of Device or Device Component
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
H3: PRODUCT ANALYSIS: PART# VPB13, LOT# 221828512 VISUAL OPTICAL VISUAL AND OPTICAL INSPECTION CONFIRMED THE CANNULA WAS RETURNED WITH THE HUB PULLED FROM THE SHAFT. ON THE CANNULA, THE KNURLING WAS PRESENT AS WELL AS THE FLAIR ON THE BACKSIDE OF THE SHAFT. IT APPEARS THE SHAFT WAS SEPARATED FROM THE PLASTIC HUB FROM EXCESSIVE FORCE. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
Additional Manufacturer Narrative
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
Description of Event or Problem
INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER (HCP) VIA MANUFACTURER REPRESENTATIVE REGARDING A PATIENT FOR UNKNOWN SPINAL THERAPY. IT WAS REPORTED THAT WHEN REMOVING CANNULA FROM PATIENT (AFTER RETRIEVING BIOPSY) THE TOP OF THE CANNULA CAME DETACHED. THERE WAS PATIENT INVOLVED IN THE EVENT AND NO FURTHER COMPLICATIONS OR SYMPTOMS WERE REPORTED. ADDITIONAL INFORMATION WAS RECEIVED VIA MANUFACTURER REPRESENTATIVE THAT THERE WERE NO FRAGMENTS LEFT INSIDE THE PATIENT. HCP WAS ABLE TO REMOVE THE DETACHED CANNULA USING A SURGICAL CLAMP. THE PRODUCT CAME IN CONTACT WITH THE PATIENT. PRE-OP DIAGNOSIS AND PROCEDURE INVOLVED IN THE EVENT WAS BONE BIOPSY.