inforMED
MalfunctionMRZ

CARTRIDGE 3ML RG MEDICATION

Received Oct 29, 2021 · Event occurred Oct 21, 2021

Report MW5105043 · MDR key 12728420

Device

Generic name

Accessories, Pump, Infusion

Lot number

210408

Product problems

  • Incorrect, Inadequate or Imprecise Result or Readings

Patient

63 YR

  • Insufficient Information
  • Insufficient Information

Narrative

Description of Event or Problem

INBOUND; PT REPORTS DRAWING UP 2.8 ML OF TREPROSTINIL IN SYRINGE AND AFTER LOADING, FILLING TUBING AND STARTING INFUSION, THE PUMP DISPLAYS HER ONLY HAVING 1.5 ML LEFT, WHICH WILL NOT LAST HER 72 HOURS. HER SYRINGE ACTUALLY HAS 2.6 MLS IN IT BUT SINCE THE PUMP CANNOT READ THAT, IT WILL THINK IT IS EMPTY SOONER. ASSISTED PT WITH RELOADING THE SYRINGE AND NO CHANGE IN VOLUME. PRIOR TO DRAWING UP DRUG, PT LUBRICATES THE SYRINGE BY PUSHING AND PULLING THE PLUNGER, TWELVE TIMES. THE LOT NUMBER FOR SYRINGE IS 210408, EXP: 4/7/2026. NO FURTHER DETAILS PROVIDED. WHELE LLC.