inforMED
InjuryPZX

ATRICLIP LAA EXCLUSION SYSTEM WITH PRELOADED GILLINOV-COSGROVE CLIP

Received Oct 20, 2021 · Event occurred Sep 9, 2021

Report 3011706110-2021-00045 · MDR key 12666869

Device

Generic name

Atriclip Laa Exclusion System With Preloaded Gillinov-cosgrove Clip

Manufacturer

Atricure, Inc.

Model number

PRO240

Catalog number

A000966

Lot number

UNKNOWN

Product problems

  • Use of Device Problem
  • Use of Device Problem

Patient

Not reported

  • Cardiac Perforation
  • Cardiac Perforation

Narrative

Additional Manufacturer Narrative

(B)(4). THE DEVICE WAS NOT RETURNED FOR EVALUATION AND A DEVICE HISTORY REVIEW WAS UNABLE TO BE COMPLETED AS THE RELEVANT LOT NUMBER FOR THE DEVICE WAS NOT REPORTED OR ABLE TO BE SUBSEQUENTLY ASCERTAINED.

Description of Event or Problem

IT WAS REPORTED THAT ON (B)(6) 2021 A PATIENT UNDERWENT AN ON-PUMP OPEN STERNOTOMY LVAD (LEFT VENTRICULAR ASSIST DEVICE) WITH LEFT ATRIAL APPENDAGE MANAGEMENT. WHEN THE PHYSICIAN WAS ATTEMPTING TO DEPLOY THE PRO240 CLIP, THE LEFT ATRIAL APPENDAGE WAS INJURED BY THE DISTAL TIP OF THE CLIP. THE SURGEON REMOVED THE PRO240 CLIP AND A SECOND CLIP, AN ACHV40 WAS PLACED WHICH MANAGED THE PERFORATION. THE LVAD WAS COMPLETED AND PATIENT WAS STABLE POST PROCEDURE. THERE WAS NO REPORTED DEVICE MALFUNCTION, AND THE ADVERSE EVENT WAS THE RESULT OF A PROCEDURAL COMPLICATION.