inforMED
MalfunctionPML

5.5 EXP VERSE FEN SCR 6.0X45

Received Oct 19, 2021 · Event occurred Sep 13, 2021

Report 1526439-2021-02204 · MDR key 12658071

Device

Generic name

Bone Cement, Posterior Screw Augmentation

Model number

199723645

Catalog number

199723645

Product problems

  • Break
  • Break

Patient

Not reported

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. WITHOUT A LOT NUMBER THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED. PRODUCT WAS NOT RETURNED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem

DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2021 DURING A POSTERIOR LUMBAR INTERBODY FUSION, FOUR (4) CAPS AND ONE (1) SCREW WERE BROKEN OFF AND ALL THE BROKEN PARTS WERE REMOVED. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT OR SURGICAL DELAY. THIS REPORT IS FOR ONE (1) EXPEDIUM VERSE SPINE SYSTEM FENESTRATED CORTICAL FIX POLYAXIAL SCREW 5.5 6.0 X 45MM. THIS IS REPORT 5 OF 5 FOR (B)(4).

Additional Manufacturer Narrative

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. DEVICE HISTORY REVIEW: THE DEVICE LOT NUMBER IS UNKNOWN, THEREFORE A MRE REVIEW COULD NOT BE PERFORMED. IF MORE INFORMATION BECOME AVAILABLE, THE RECORD WILL BE RE-ASSESSED THE DEVICE WAS NOT RETURNED. A PHOTO-INVESTIGATION WAS PERFORMED ON THE IMAGES UPON INSPECTING ALL THE IMAGES PROVIDED UNDER NOTES & ATTACHMENTS SECTION, IT IS OBSERVED 5.5 EXP VERSE FEN SCR 6.0X45 IS BROKEN AT THE THREADS, HENCE CONFIRMING THE ALLEGATION. AS THE DEVICE WAS NOT RETURNED, AN AS-RECEIVED CONDITION COULD NOT BE ASSESSED, AND A DIMENSIONAL INSPECTION AND DOCUMENT/SPECIFICATION REVIEW WERE NOT COMPLETED. CONCLUSION THE COMPLAINT CONDITION CAN BE CONFIRMED DURING PHOTO/VIDEO INVESTIGATION. DURING THE INVESTIGATION, NO PRODUCT DESIGN ISSUES OR DISCREPANCIES WERE OBSERVED. NO MANUFACTURING ISSUES WERE NOTED DURING INVESTIGATION. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST MARKET SAFETY SURVEILLANCE ACTIVITIES. THEREFORE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTIVE ACTION IS PROPOSED. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.