COMPAX 40E
Received Dec 16, 2008 · Event occurred Oct 22, 2008
Report 2126677-2008-00104 · MDR key 1265724
Device
Manufacturer
Ge Medical Systems, LlcProduct problems
- Unintended System Motion
- Device Inoperable
Patient
Not reported
- No Known Impact Or Consequence To Patient
Narrative
Description of Event or Problem
IT WAS REPORTED THAT THE TABLE LOCKS DID NOT ACTUATE WHEN THE FOOT PEDAL WAS RELEASED AS THE PT WAS BEING PREPARED FOR AN EXAM, CAUSING UNEXPECTED LATERAL MOTION OF THE TABLETOP WITHOUT RESISTANCE (LATERAL FLOAT). THERE WAS NO INJURY REPORTED. THIS SITUATION COULD CONTRIBUTE TO AN INJURY IF A PATIENT OR OPERATOR WERE UNAWARE OF THIS CONDITION WHILE LOADING A PT. THE ENSUING INSTABILITY COULD LEAD TO A FALL.
Additional Manufacturer Narrative
UPON DISCOVERY OF THE UNEXPECTED LATERAL FLOAT, THE SITE WAS STILL ABLE TO USE THE TABLE FOR THE EXAM WHEN THEY USED THE REDUNDANT MECHANICAL INTERLOCK TO INHIBIT LATERAL MOTION. THE GE FIELD ENGINEER (FE) EVALUATED THE SYSTEM AND FOUND A DAMAGED MICROSWITCH. THE FE REPLACED THE MICROSWITCH AND VERIFIED THAT THE TABLE LOCKS WERE PERFORMING ACCORDING TO SPECIFICATIONS.