IFUSE IMPLANT SYSTEM

Received Oct 14, 2021 · Event occurred Sep 16, 2021

Report 3007700286-2021-00102 · MDR key 12637305

Device

Generic name

Orthopedic Rod

Manufacturer

Si-bone, Inc.

Model number

7050M-90

Lot number

493763

Product problems

  • Malposition of Device
  • Malposition of Device

Patient

54 YR

  • Undesired Nerve Stimulation
  • Undesired Nerve Stimulation

Narrative

Additional Manufacturer Narrative

BASED ON THE INFORMATION PROVIDED, REVIEW OF THE SURGICAL TECHNIQUE MANUAL, IFU, CERTIFICATES OF CONFORMANCE AND FMEA, THERE IS NO INDICATION OF DEVICE FAILURE AND NO INDICATION THAT THE DEVICE WAS OUT OF SPECIFICATION. THE MOST PROBABLE ROOT CAUSE IS USER ERROR: USING TOO LONG OF AN IMPLANT OR INSTALLING THE IMPLANT TOO DEEP. ADDITIONALLY, PER (B)(4) (US), SI-BONE SURGICAL TECHNIQUE MANUAL (C-ARM AND O-ARM), THE SURGEON IS INSTRUCTED AS FOLLOWS: "PRIOR TO CLOSURE, ALWAYS OBTAIN FINAL FLUOROSCOPIC IMAGES IN THE LATERAL, INLET, AND OUTLET VIEWS TO CONFIRM NO CORTICAL WALL BREACH, FORAMEN BREACH, OR OTHER MALPOSITION."

Description of Event or Problem

THE PATIENT HAD RIGHT SIDE SI JOINT ARTHRODESIS IN (B)(6) 2016 WHERE THREE IMPLANTS WERE INSTALLED. THE PATIENT LATER REPORTED A RECURRENCE OF PAIN SYMPTOMS. THE NEW SURGEON DETERMINED BASED ON THE IMAGING REPORTS THAT THE MIDDLE-POSITIONED IMPLANT MAY HAVE BEEN IMPINGING ON THE NEUROFORAMEN. (B)(6) 2021, THE SURGEON PERFORMED A REVISION PROCEDURE WHERE HE REMOVED THE MIDDLE POSITIONED IFUSE IMPLANT USING THE IFUSE REMOVAL ADAPTER AS IT WAS SOLIDLY FIXED IN BONE. THE IMPLANT WAS REPLACED WITH A SHORTER IMPLANT. NO OTHER PREEXISTING IMPLANTS WERE ADJUSTED OR REMOVED. THE STATUS OF THE PATIENT FOLLOWING THE REVISION PROCEDURE IS NOT KNOWN.