OES CHOLEDOCHO FIBERSCOPE

Received Oct 9, 2021

Report 8010047-2021-12945 · MDR key 12602550

Device

Generic name

Choledocho Fiberscope

Model number

CHF-P60

Product problems

  • Mechanical Problem
  • Mechanical Problem

Patient

Not reported

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE SUBJECT DEVICE WAS NOT RETURNED TO OMSC FOR EVALUATION BUT WAS RETURNED TO SORC. SORC CHECKED THE SUBJECT DEVICE AND FOUND THE REPORTED PHENOMENON. OMSC REVIEWED THE MANUFACTURE HISTORY (DHR) OF THE SUBJECT DEVICE AND CONFIRMED NO IRREGULARITY. BASED UPON THE INFORMATION FROM SORC THAT SORC FOUND THE PIN HOLE ON THE BENDING RUBBER DURING THE INSPECTION, OMSC SURMISED THAT THE REPORTED PHENOMENON WAS ATTRIBUTED TO THE ACTION FAILURE OF THE ANGULATION DUE TO THE CORROSION OF THE WIRE AND/OR THE BENDING SECTION WHICH MIGHT BE CAUSED BY THE WATER INGRESS FROM THE PIN HOLE ON THE BENDING RUBBER. IF ADDITIONAL INFORMATION IS RECEIVED, THIS REPORT WILL BE SUPPLEMENTED

Description of Event or Problem

OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED THAT DURING THE INCOMING INSPECTION FOR REPAIR AT OLYMPUS SERVICE OPERATION REPAIR CENTER (SORC), IT WAS FOUND THAT THE ANGULATION WAS LOCKED DUE TO CORROSION OF THE ANGULATION MECHANISM. THE OCCURRENCE DATE OF THE EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT INJURY ASSOCIATED WITH THE EVENT.