MalfunctionFBO

PANOVIEW

Received Oct 7, 2021 · Event occurred Sep 21, 2021

Report 12589862 · MDR key 12589862

Device

Generic name

Cystourethroscope

Manufacturer

Richard Wolf Gmbh

Model number

8626431

Catalog number

8626.431

Product problems

  • Component Missing

Patient

Not reported

  • Foreign Body In Patient
  • Foreign Body In Patient

Narrative

Description of Event or Problem

PHYSICIAN WAS DOING A CYSTOSCOPY ON THE PATIENT WITH A RICHARD WOLF GMBH PANOVIEW CYSTOSCOPE. PHYSICIAN NOTICED A FOREIGN BODY INSIDE THE BLADDER OF THE PATIENT. MEDICAL STAFF CHECKED AND THE SCOPE WAS MISSING ITS LENS COVER. THE SURGEON RETRIEVED THE LENS COVER BY USING A GRASPER. NO HARM TO THE PATIENT.