inforMED
OtherOAO

AXP AUTOXPRESS PLATFORM

Received Dec 8, 2008 · Event occurred Nov 11, 2008

Report 2950374-2008-00003 · MDR key 1258146

Device

Model number

8-5101

Catalog number

8-5101

Lot number

27066-0000-02

Product problems

  • Device Contamination with Chemical or Other Material

Patient

Not reported

  • No Information

Narrative

Description of Event or Problem

A CUSTOMER CALL CAME IN ON (B)(6)2008, TO REPORT THAT WHITE DEBRIS WAS FOUND IN THE DISPOSABLE FREEZING BAG.

Additional Manufacturer Narrative

WE HAVE COMPLETED OUR EVAL OF THE RETURNED DISPOSABLE BAGSET. THE WHITE DEBRIS WAS SENT OUT FOR FOURIER TRANSFORM INFRARED (FTIR) SPECTROSCOPY FOR IDENTIFICATION. THE PARTICLE WAS IDENTIFIED AS BAGSET STOPCOCK MATERIAL. THIS IS AN ISOLATED EVENT FOR THIS LOT OF DISPOSABLE BAGSETS FROM THIS MFR. WE WILL CONTINUE TO MONITOR AND TREND. NO ADVERSE EVENTS HAVE BEEN REPORTED FOR THIS OCCURRENCE.