UNK - PLATES: 4.5 MM LCP CONDYLAR PLATE
Received Sep 24, 2021 · Event occurred Sep 30, 2020
Report 8030965-2021-08052 · MDR key 12529844
Device
Generic name
Condylar Plate Fixation Implant
Manufacturer
Synthes GmbhProduct problems
- Break
- Break
Patient
Not reported
- Nonunion/Delayed-union Bone Fracture
- Nonunion/Delayed-union Bone Fracture
Narrative
Additional Manufacturer Narrative
EXACT DATE OF EVENT IS UNKNOWN; SEPTEMBER 30, 2020 IS THE DATE THE LITERATURE ARTICLE WAS PUBLISHED. THIS REPORT IS FOR AN UNKNOWN LCP CONDYLAR PLATE/ UNKNOWN LOT. PART AND LOT NUMBER ARE UNKNOWN; UDI NUMBER IS UNKNOWN. COMPLAINANT PART IS EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/ INVESTIGATION, BUT HAS YET TO BE RECEIVED. WITHOUT A LOT NUMBER THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED. PRODUCT WAS NOT RETURNED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/ OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
Description of Event or Problem
THIS REPORT IS BEING FILED AFTER THE REVIEW OF THE FOLLOWING JOURNAL ARTICLE: JONES B. M., NUGENT L., GRIFFIN S. J., (2020) FEMORAL CONDYLAR LOCKING COMPRESSION PLATES: A CASE SERIES. THE RURAL DISTRICT GENERAL HOSPITAL EXPERIENCE, JOURNAL OF SURGICAL CASE REPORTS, VOLUME 9, PAGES 1¿5 (UNITED KINGDOM), DOI: 10.1093/JSCR/RJAA380. THIS STUDY CASE SERIES AIMS TO EXAMINE OUTCOMES FOR PATIENTS MANAGED IN A RURAL SETTING. RECENT INSERTIONS OF CONDYLAR LOCKING COMPRESSION PLATES (LCPS) AT A RURAL DISTRICT GENERAL HOSPITAL WERE EXAMINED. BETWEEN 2014 AND 2019, 50 MOST RECENT DISTAL FEMORAL FRACTURES COMPRISING OF 40 WOMEN AND 10 MEN, WITH MEAN AGE AT PRESENTATION OF 77.38 YEARS TREATED WITH THE SYNTHES 4.5/5.0 LCP CONDYLAR PLATE OPERATED AT YSBYTY GWYNEDD DISTRICT GENERAL HOSPITAL (DGH) (NORTH WALES) WERE ASSESSED, PRIOR TO 6 MONTHS FROM COMMENCING THE STUDY TO ALLOW FOR ADEQUATE FOLLOW-UP. AVERAGE LENGTH OF FOLLOW-UP WAS 43 WEEKS. THE FOLLOWING COMPLICATIONS WERE REPORTED AS FOLLOWS: CASE 3, ANOTHER CASE OF FATIGUE FRACTURE OF THE PLATE PRIOR TO BONY UNION. HERE CERCLAGE WIRES CLOSE TO THE FRACTURE SITE MAY HAVE HAD THE SAME EFFECT AS PLACING SCREWS IMMEDIATELY ADJACENT TO THE FRACTURE. PLATE FATIGUE FRACTURE PRIOR TO SUCCESSFUL UNION (FIG 7) THIS REPORT IS FOR AN UNKNOWN LCP CONDYLAR PLATE. THIS IS REPORT 4 OF 8 FOR (B)(4).
Additional Manufacturer Narrative
PRODUCT COMPLAINT # (B)(4). DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H10 ADDITIONAL NARRATIVE: D4: UPDATED IMPACTED PRODUCT H3, H6: THE INVESTIGATION COULD NOT BE COMPLETED, NO PRODUCT WAS RECEIVED; NO CONCLUSION COULD BE DRAWN AT THE TIME OF FILING THIS REPORT. THE PRODUCT WAS NOT RETURNED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.