ARCHITECT CEA REAGENT KIT
Received Sep 7, 2021 · Event occurred Aug 13, 2021
Report 3008344661-2021-00164 · MDR key 12433509
Device
Generic name
System, Test, Carcinoembryonic Antigen
Manufacturer
Abbott IrelandModel number
07K68-27Catalog number
07K68-27
Lot number
26467FN00
Product problems
- Low Test Results
- Low Test Results
Patient
71 YR
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE COMPLAINT INVESTIGATION INCLUDED A REVIEW OF DATA AND INFORMATION PROVIDED BY THE CUSTOMER, SEARCH FOR SIMILAR COMPLAINTS, TICKET TRENDING REVIEW, LABELING REVIEW, DEVICE HISTORY RECORD REVIEW AND FIELD DATA REVIEW. RETURN TESTING WAS NOT COMPLETED AS RETURNS WERE NOT AVAILABLE. TRENDING REVIEW DID NOT IDENTIFY ANY TRENDS FOR THE LIKELY CAUSE LOT OR COMPLAINT ISSUE. DEVICE HISTORY RECORD REVIEW DID NOT IDENTIFY ANY NON-CONFORMANCES OR DEVIATIONS FOR THE LIKELY CAUSE LOT. THE OVERALL PERFORMANCE OF ARCHITECT CEA REAGENT IN THE FIELD WAS REVIEWED USING DATA GATHERED FROM CUSTOMERS WORLDWIDE. THE MEDIAN PATIENT POPULATION RESULT FOR LOT 26467FN00 IS COMPARABLE WITH ALL OTHER LOTS IN THE FIELD AND WITHIN ESTABLISHED BASELINES. LABELING WAS REVIEWED AND SUFFICIENTLY ADDRESSES THE CUSTOMER¿S ISSUE. NO SYSTEMIC ISSUE OR DEFICIENCY WITH THE ARCHITECT CEA REAGENT LOT 26467FN00 WAS IDENTIFIED.
Additional Manufacturer Narrative
PATIENT IDENTIFIER: (B)(6). AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.
Description of Event or Problem
THE CUSTOMER OBSERVED FALSELY DEPRESSED ARCHITECT CEA RESULTS GENERATED ON AN ARCHITECT I2000 PROCESSING MODULE ON A (B)(6) MALE UNDERGOING RADIATION THERAPY FOR LUNG CANCER. THE FOLLOWING RESULTS WERE PROVIDED. SID (B)(6) INITIAL RESULT = 1.47 NG/ML, REPEAT RESULT = 32.63 NG/ML. THE PATIENT HAD BEEN TRENDING AROUND 30 NG/ML. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.