DYNAMIC INLET, POWER EXPRESS
Received Sep 1, 2021 · Event occurred Aug 9, 2021
Report 2050012-2021-00057 · MDR key 12411371
Device
Generic name
Radioassay, Vitamin B12
Manufacturer
Beckman CoulterModel number
DYNAMIC INLET, POWER EXPRESSCatalog number
PK201605005
Product problems
- Imprecise or Erratic Results
- Imprecise or Erratic Results
Patient
Not reported
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
THE CUSTOMER REPORTED THAT A STAT RACK IN BAY 7 OF DYNAMIC INLET WAS MISIDENTIFIED AS CENTRIFUGE BYPASS RACK IN POWER EXPRESS SYSTEM. AS A RESULT, THREE SAMPLE TUBES WERE ROUTED ON THE LINE AND WERE SORTED TO THE CONNECTED INSTRUMENTS WHICH GENERATED RESULTS, VALIDATED, AND HOSTED TO LIS (LABORATORY INFORMATION SYSTEM). THE UNCENTRIFUGED SAMPLE ROUTED TO DXC 800 INSTRUMENT. THE CUSTOMER REPORTED RECEIVING ERRONEOUS PATIENT RESULTS FOR MULTIPLE ANALYTES FOR THE CMP (COMPREHENSIVE METABOLIC PROFILE). THE SAMPLE WAS WHOLE BLOOD AND AN ERRONEOUS RESULTS WERE GENERATED FOR NA (SODIUM), K (POTASSIUM), CL (CHLORIDE), CO2 (CARBON DIOXIDE), BUN (BLOOD UREA NITROGEN), GLU (GLUCOSE), CREA (CREATININE), CALC (CALCIUM), TP (TOTAL PROTEIN), ALB (ALBUMIN), TBIL (TOTAL BILIRUBIN), TRIG (TRIGLYCERIDES), CHOLESTEROL ALP (ALKALINE PHOSPHATASE), ALT (ALANINE AMINOTRANSFERASE) AND AST (ASPARTATE AMINOTRANSFERASE). THE ERRONEOUS RESULTS WERE REPORTED OUTSIDE OF THE LAB. THERE WAS NO CHANGE IN PATIENT TREATMENT IN ASSOCIATION WITH THIS EVENT.
Additional Manufacturer Narrative
A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED AND EVALUATED THE DEVICE. THE FIELD SERVICE ENGINEER (FSE) REPLACED THE RACK SWITCHES TO RESOLVE THE ISSUE. (B)(4). NO PATIENT DEMOGRAPHIC INFORMATION (AGE, GENDER, WEIGHT, RACE OR ETHNICITY) WAS PROVIDED.