MalfunctionFSY
CHROMOPHARE OR LIGHT CEILING-MOUNTED
Received Aug 31, 2021 · Event occurred Aug 2, 2021
Report 0008010153-2021-00007 · MDR key 12397980
Device
Generic name
Light, Surgical, Ceiling Mounted
Manufacturer
Berchtold Gmbh & Co. KgModel number
CH00000001Catalog number
CH00000001
Product problems
- Mechanical Problem
- Mechanical Problem
Patient
Not reported
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
IT WAS REPORTED THE BLUE DOME SEPARATED FROM THE LIGHT HEAD OUTSIDE OF A CASE. THERE WAS NO REPORTED PATIENT INVOLVEMENT, INJURIES, OR ADVERSE CONSEQUENCES. THE ROOT CAUSE OF THIS IS DUE TO THE LACK OF ADHESION BETWEEN THE BLUE DOME AND THE PLASTIC PORTION OF THE LIGHT HEAD COVER.
Description of Event or Problem
IT WAS REPORTED THE OR 1 BLUE DOME COVER COMING OFF OF LIGHT HEAD. THERE WERE NO REPORTED INJURIES OR ADVERSE CONSEQUENCES.