inforMED
MalfunctionFSY

CHROMOPHARE OR LIGHT CEILING-MOUNTED

Received Aug 31, 2021 · Event occurred Aug 4, 2021

Report 0008010153-2021-00006 · MDR key 12397748

Device

Generic name

Light, Surgical, Ceiling Mounted

Model number

CH00000001

Catalog number

CH00000001

Product problems

  • Mechanical Problem
  • Mechanical Problem

Patient

Not reported

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THE BLUE DOME FELL OFF OF THE F-GEN LIGHT HEAD IN OR 1. THERE WERE NO REPORTED INJURIES OR ADVERSE CONSEQUENCES.

Additional Manufacturer Narrative

IT WAS REPORTED THAT THE BLUE DOME SEPARATED FROM AN F GENERATION LIGHT. THE LIGHT HEAD REPORTED IS WITHIN THE SCOPE OF (B)(4) AND WILL BE HANDLED ACCORDINGLY. THERE WAS NO REPORTED PATIENT INVOLVEMENT, NO INJURIES, NO ADVERSE CONSEQUENCES OR DELAY. PER (B)(4), THE SEVERITY IS S0.