HYGENIC DENTAL DAM THIN GREEN
Received Aug 27, 2021 · Event occurred Mar 29, 2021
Report 2416455-2021-00004 · MDR key 12377770
Device
Product problems
- Material Split, Cut or Torn
- Material Split, Cut or Torn
Patient
Not reported
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
INVESTIGATION OF RETURNED PRODUCT: TYPICAL VALUES FOR FRESHLY PRODUCED DENTAL DAM ARE TENSILE GREATER THAN OR EQUAL TO 3,500 LBF/IN² AND ULTIMATE ELONGATION GREATER THAN OR EQUAL TO 700% WITH REFERENCE TO TYPICAL MODULUS AT 100% ELONGATION AT 120 LBF/IN². PROPERTIES WILL DETERIORATE AS THE DAM AGES. THE MEAN VALUE FOR TENSILE IS 4,414 LBF/IN² WHICH IS WITHIN SPECIFICATION FOR NEWLY PRODUCED DENTAL DAM AS WELL AS ULTIMATE ELONGATION AT 788% WITH 110 LBF/IN² AS THE MODULUS AT 100% LBF/IN². FROM THE DATA PRODUCED, IT IS DETERMINED THAT THE DENTAL DAM IS AS EXPECTED AND NO ABNORMAL RESULTS WERE FOUND. THE FIT, FORM AND FUNCTION OF THE DAM IS NOT COMPROMISED. A REVIEW OF PAST COMPLAINTS WAS PERFORMED REGARDING TEARS FOR THIS SHEETING LOT FROM MAY 2019 UNTIL PRESENT. THIS IS THE FIRST COMPLAINT RECORDED FOR THIS LOT/PART; THEREFORE, NO RECURRENCE WAS IDENTIFIED.
Description of Event or Problem
THE CUSTOMER INITIALLY REPORTED THAT THE DENTAL DAM TEARS VERY QUICKLY. IT WAS CONFIRMED ON (B)(6) 2021 THAT THE DENTAL DAM TORE DURING TREATMENT AND THAT THE ROOT CANAL TREATMENT HAD TO BE RESTARTED. THE CUSTOMER WAS CONTACTED ON (B)(6) 2021 TO OBTAIN ADDITIONAL INFORMATION REGARDING ANY OTHER MEDICAL DEVICES THAT WERE BEING USED AT THE TIME OF THE INCIDENT AND TO DETERMINE THE NUMBER OF PATIENTS THAT WERE AFFECTED. DURING THIS FOLLOW UP, IT WAS DETERMINED THAT AN ADVERSE EVENT DID NOT ACTUALLY OCCUR. THE DENTIST INDICATED THAT THE DENTAL DAM TORE PRIOR TO PROCEDURE AND WAS JUST PROVIDING AN EXAMPLE OF WHAT COULD HAVE HAPPENED IF THE DENTAL DAM WOULD HAVE TORN DURING PROCEDURE (RESTART THE ROOT CANAL TREATMENT).
Description of Event or Problem
THE CUSTOMER INITIALLY REPORTED THAT THE DENTAL DAM TEARS VERY QUICKLY. IT WAS CONFIRMED ON 07/08/2021 THAT THE DENTAL DAM TORE DURING TREATMENT AND THAT THE ROOT CANAL TREATMENT HAD TO BE RESTARTED.