Device

Generic name

Infant Heel Warmer (chemical Heat Pack)

Manufacturer

Mo016 Mo-moberly

Model number

11460-010T

Catalog number

11460-010T

Lot number

V1E144

Product problems

  • Burst Container or Vessel
  • Burst Container or Vessel

Patient

63 YR

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

A REVIEW OF THE DEVICE HISTORY RECORD WAS CONDUCTED FOR THE REPORTED LOT NUMBER AND REVEALED IT WAS MANUFACTURED AND RELEASED ON 5/20/2021 TO PREDETERMINED SPECIFICATIONS. NO ANOMALIES WERE FOUND DURING REVIEW OF THE RECORDS. THE SAMPLES RECEIVED FOR INVESTIGATION WERE ALREADY ACTIVATED. THE EXACT ROOT CAUSE CANNOT BE DETERMINED BASED ON THE SAMPLES RECEIVED. A WARNING TO THE LABEL FOR THE CUSTOMER TO ACTIVATE AWAY FROM ALL PEOPLE WILL BE ADDED TO THE DEVICE. ADDITIONALLY, WE WILL CONTINUE TO MONITOR COMPLAINT TRENDS FOR THIS TYPE OF INCIDENT.

Description of Event or Problem

CUSTOMER REPORT WE ARE CONTINUING HAVING ISSUES WITH THE HEEL WARMERS. THE HEEL WARMERS ARE ALREADY ACTIVATED PRIOR TO USE. I HAD 6 MORE BROUGHT INTO MY OFFICE AND ONE REPORTEDLY BURST AT THE SEAM WHEN A NURSE ACTIVATED IT.