COMPAX 40E
Received Nov 12, 2008 · Event occurred Oct 24, 2008
Report 2126677-2008-00092 · MDR key 1235628
Device
Manufacturer
Ge Medical Systems, LlcProduct problems
- Unintended System Motion
Patient
Not reported
- No Known Impact Or Consequence To Patient
Narrative
Description of Event or Problem
IT WAS REPORTED THE TABLE LOCKS DID NOT ACTUATE, CAUSING THE TABLETOP TO UNEXPECTEDLY MOVE IN THE LONGITUDINAL AND LATERAL DIRECTIONS WITHOUT RESISTANCE (FREE FLOAT). THERE WAS NO INJURY REPORTED. THIS SITUATION COULD CONTRIBUTE TO AN INJURY IF A PATIENT OR OPERATOR WERE UNAWARE OF THIS CONDITION WHILE LOADING OR UNLOADING A PATIENT. THE ENSUING INSTABILITY COULD LEAD TO A FALL.
Additional Manufacturer Narrative
THE GE FIELD ENGINEER (FE) REPORTED THAT THE TABLE LOCKS DID NOT ACTUATE DUE TO LOSS OF POWER. THE FE REPLACED THE TABLE FUSE, AND THE FLOAT COULD NOT BE DUPLICATED. THE FE VERIFIED THAT THE TABLE LOCKS ARE PERFORMING ACCORDING TO SPECIFICATIONS.