OPMI PICO

Received Aug 19, 2021 · Event occurred Jul 6, 2021

Report 9615010-2021-00003 · MDR key 12345882

Device

Generic name

Opmi Pico

Product problems

  • Detachment of Device or Device Component
  • Detachment of Device or Device Component

Patient

Not reported

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE ISSUE WAS CAUSED DUE TO A LOOSE SAFETY SCREW ON THE OPMI. THE SERVICE MANUAL CLEARLY STATES THAT THE SCREW MUST BE TIGHTENED FIRMLY. THE MANUFACTURING, DESIGN, AND LABELING, IS SUFFICIENT TO PREVENT SUCH AN ISSUE. THE CUSTOMER DECLINED A DEVICE REPAIR AND EVALUATION BY THE MANUFACTURER AND CLAIMED TO HAVE REPAIRED THE DEVICE BY THEIR OWN CLINICAL ENGINEERING DEPARTMENT. IT IS NOT KNOWN WHO IS RESPONSIBLE FOR THE LOOSE SCREW. HOWEVER, DUE TO THE FACT THAT THE LAST PREVENTIVE MAINTENANCE BY ZEISS WAS PERFORMED ONE YEAR AGO WITH NO DEVIATION, AND THE CUSTOMER REPAIRED THE DEVICE BY THEIR OWN ENGINEERING DEPARTMENT, THIS ISSUE WAS POSSIBLY CAUSED BY IMPROPER HANDLING/UNAUTHORIZED PERSON TAMPERING WITH THE DEVICE.