Received Aug 19, 2021 · Event occurred Jun 18, 2021

Report 2134070-2021-00018 · MDR key 12344626

Device

Generic name

Bit, Drill

Manufacturer

Sterilmed, Inc.

Model number

KOMKM166-00-01

Catalog number

KOMKM166-00-01

Product problems

  • Detachment of Device or Device Component
  • Detachment of Device or Device Component

Patient

64 YR

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE DEVICE HAS NOT YET BEEN RETURNED FOR ANALYSIS. THE LOT NUMBER IS UNKNOWN, IT IS NOTED THAT FOR THIS ACCOUNT, THE DRILL BIT IS TAKEN OUT OF PACKAGING IN STERILE PROCESSING DEPARTMENT AND STERILIZED BEFORE BEING SENT TO THE OPERATING ROOM. THIS COMPLAINT WAS CREATED BASED OFF OF INFORMATION RECEIVED FROM A USER FACILITY MEDWATCH REPORT # (B)(4). IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. MANUFACTURER'S REF. NO: (B)(4).

Description of Event or Problem

IT WAS REPORTED THAT A (B)(6) FEMALE PATIENT UNDERWENT AN ARTHROPLASTY PROCEDURE WITH A REPROCESSED DRILL BIT TWIST STAINLESS STEEL AND A SMALL PIECE BROKE OFF AND LODGED IN THE ANTERIOR ASPECT OF THE HUMERUS. THE FRAGMENT WAS LEFT IN PLACE INTENTIONALLY SO AS NOT TO COMPROMISE THE INTEGRITY OF THE PROXIMAL HUMERUS STEM FOR THE PROSTHESIS. THERE WAS NO DIFFICULTY DURING THE PROCEDURE OR PATIENT CIRCUMSTANCE THAT CONTRIBUTED TO THE BREAK. THIS EVENT HAS BEEN ASSESSED AS A REPORTABLE EVENT.

Additional Manufacturer Narrative

IT WAS REPORTED THAT A 64-YEAR-OLD, FEMALE PATIENT UNDERWENT AN ARTHROPLASTY PROCEDURE WITH A REPROCESSED DRILL BIT TWIST STAINLESS STEEL AND A SMALL PIECE BROKE OFF AND LODGED IN THE ANTERIOR ASPECT OF THE HUMERUS. THE DEVICE WAS RETURNED ON 23-AUG-2021. THE RETURNED DEVICE WAS OBSERVED TO BE BROKEN ABOUT 4MM ABOVE THE PROXIMAL BASE OF THE THREADING. THE BREAK WAS JAGGED AND BIOLOGICAL CONTAMINANTS REMAINED ON THE SURFACE. THE JAGGED EDGE IS INDICATIVE OF EXCESSIVE PRESSURE BEING APPLIED TO THIS INSTRUMENT, BENDING THE MATERIAL, LIKELY CAUSING THE DEVICE TO STRUCTURALLY FAIL AND SPLIT APART. THE REMAINING CHARACTERISTICS OF THE DEVICE, INCLUDING ITS 2.0MM DIAMETER, ARE CONSISTENT WITH BEING PRODUCT CODE KM166-00-01, WHICH IS MADE OF MEDICAL-GRADE STAINLESS STEEL. THE SURFACE WAS EXAMINED UNDER A MAGNIFICATION SCOPE, AND NO REPROCESSING MARK WAS OBSERVED. PER THE INSPECTION AND TESTING PROCEDURE FOR SHARP DEVICE, FOR DEVICES FULLY MADE OUT OF METAL I.E. DRILL BITS, SAW BLADES, ETC., A DREMEL ENGRAVER IS USED TO ETCH A MARK ON THE PROXIMAL END OF THE DEVICE. AS THERE WAS NO PROVIDED LOT NUMBER AND THERE IS NO OBSERVED REPROCESSING MARK, IT CANNOT BE VERIFIED THAT THIS IS A STERILMED REPROCESSED DEVICE. WHILE THIS RETURNED DEVICE IS OBSERVED AS BROKEN, AS THIS CANNOT BE CONFIRMED TO BE A STERILMED REPROCESSED DEVICE, THE REPORTED ISSUE IS NOT CONFIRMED. NO PACKAGING OR LABELING HAD BEEN RETURNED WITH THIS DEVICE; THEREFORE, THE LOT NUMBER WAS UNABLE TO BE OBTAINED. CONSEQUENTLY, NO MANUFACTURING RECORD EVALUATION WAS CONDUCTED SINCE NO LOT NUMBER WAS PROVIDED. IF ADDITIONAL INFORMATION IS RECEIVED REGARDING THIS EVENT, A SUPPLEMENTAL 3500A REPORT WILL BE SUBMITTED TO THE FDA. MANUFACTURER'S REF. NO: (B)(4).