MIS TI CFX FEN POLY 7X45
Received Aug 17, 2021 · Event occurred Jul 13, 2021
Report 1526439-2021-01715 · MDR key 12331328
Device
Generic name
Bone Cement, Posterior Screw Augmentation
Manufacturer
Medos International Sã rl ChModel number
186727745Catalog number
186727745
Lot number
295704
Product problems
- Break
- Break
Patient
Not reported
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H3, H4, H6: THE DHR OF PRODUCT CODE: 186727745. LOT : 295704. WAS ELECTRONICALLY REVIEWED AND NO NON-CONFORMANCES WERE OBSERVED DURING THE MANUFACTURING PROCESS. THE PRODUCT WAS RELEASED ON: DECEMBER 15, 2020. VISUAL INSPECTION: MIS TI CFX FEN POLY 7X45 (P/N: 186727745, LOT #: 295704) WAS RETURNED AND RECEIVED AT US CUSTOMER QUALITY (CQ). UPON VISUAL INSPECTION, THE FLEX BALL COMPONENT OF THE DEVICE OBSERVED TO BE BROKEN, WHICH IN TURN CONTRIBUTED FOR THE SHANK TO DISASSEMBLE FROM THE HEAD. NO OTHER ISSUES WERE OBSERVED WITH THE COMPLAINT DEVICE. DEVICE FAILURE/DEFECT IDENTIFIED? YES. DIMENSIONAL INSPECTION: COMPLAINT RELEVANT DIMENSIONAL ANALYSIS CANNOT BE PERFORMED DUE TO THE POST-MANUFACTURING DAMAGE AND DESIGN OF THE DEVICE. DOCUMENT/SPECIFICATION REVIEW: BASED ON THE MANUFACTURED DATE OF THE DEVICE, THE FOLLOWING REVISIONS OF CURRENT AND MANUFACTURED DRAWINGS WERE REVIEWED: EXPEDIUM 5.5 TI, CORTICAL FIX POLYAXIAL SCREW ASSY, BOTTOM LOADING SHANK NO DESIGN ISSUES OR DISCREPANCIES WERE IDENTIFIED. COMPLAINT CONFIRMED? YES. INVESTIGATION CONCLUSION: THE COMPLAINT COULD BE CONFIRMED FOR THE RETURNED DEVICE. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE COMPLAINT CONDITION AND HENCE THE ROOT CAUSE CANNOT BE DETERMINED. THE POTENTIAL CAUSE COULD BE DUE TO UNINTENDED FORCES APPLIED ON THE DEVICE. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
Additional Manufacturer Narrative
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. THE COMPLAINT DEVICE WAS NOT RECEIVED FOR INVESTIGATION. A PHOTO INVESTIGATION WAS PERFORMEDTHE IMAGES WERE REVIEWED AND THE COMPLAINT CONDITION CAN BE CONFIRMED. THE CORTICAL FIX FENESTRATED SCREW IS BROKEN. THERE WERE NO NON-CONFORMANCES OBSERVED IN THE DHR REVIEW. A DEFINITIVE ASSIGNABLE ROOT CAUSE COULD NOT BE DETERMINED BASED ON THE PROVIDED INFORMATION. AS THE DEVICE WAS NOT RETURNED, AN AS-RECEIVED CONDITION COULD NOT BE ASSESSED, AND A DIMENSIONAL INSPECTION AND DOCUMENT/SPECIFICATION REVIEW WERE NOT COMPLETED DURING THE INVESTIGATION NO PRODUCT DESIGN ISSUES OR DISCREPANCIES WERE OBSERVED (BASED ON THE IMAGES) THAT MAY HAVE CONTRIBUTED TO THE COMPLAINT CONDITION. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE HISTORY LOT: THE DHR OF PRODUCT CODE: 186727745 LOT: 295704 WAS ELECTRONICALLY REVIEWED AND NO NON-CONFORMANCES WERE OBSERVED DURING THE MANUFACTURING PROCESS. THE PRODUCT WAS RELEASED ON: 15.12.2020 DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
Additional Manufacturer Narrative
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H10: ADDITIONAL NARRATIVE: H4 D9, H3, H6: THE DEVICE WAS RECEIVED, THE INVESTIGATION COULD NOT BE COMPLETED, AND NO CONCLUSION COULD BE DRAWN, AS PRODUCT IS ENTERING THE COMPLAINT SYSTEM. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
Additional Manufacturer Narrative
ADDITIONAL PROCODE: MNI. SUMMARY: THE COMPLAINT DEVICE WAS NOT RECEIVED FOR INVESTIGATION. A PHOTO INVESTIGATION WAS PERFORMED BASED ON THE RECEIVED IMAGES. THE IMAGES WERE REVIEWED, AND THE COMPLAINT CONDITION CAN BE CONFIRMED. THE CORTICAL FIX FENESTRATED SCREW IS BROKEN. A DEFINITIVE ASSIGNABLE ROOT CAUSE COULD NOT BE DETERMINED BASED ON THE PROVIDED INFORMATION. A MANUFACTURING RECORD EVALUATION COULD NOT BE PERFORMED AS A VALID LOT NUMBER COULD NOT BE DETERMINED FROM THE IMAGES PROVIDED. AS THE DEVICE WAS NOT RETURNED, AN AS-RECEIVED CONDITION COULD NOT BE ASSESSED, AND A DIMENSIONAL INSPECTION AND DOCUMENT/SPECIFICATION REVIEW WERE NOT COMPLETED DURING THE INVESTIGATION, NO PRODUCT DESIGN ISSUES OR DISCREPANCIES WERE OBSERVED THAT MAY HAVE CONTRIBUTED TO THE COMPLAINT CONDITION. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
Description of Event or Problem
DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT DURING ROBOT-ASSISTED LUMBAR FUSION AND INTERNAL FIXATION ON (B)(6) 2021, THE SCREW BROKE WHILE BEING INSERTED. ANOTHER SCREW WAS USED, BUT THE SAME PROBLEM OCCURRED. A THIRD DEVICE WAS USED TO COMPLETE THE SURGERY. ALL PIECES WERE REMOVED FROM THE PATIENT. PROCEDURE WAS COMPLETED WITH NO DELAY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. THIS REPORT IS FOR A MIS TI CFX FEN POLY 7X45. THIS IS REPORT 1 OF 2 FOR (B)(4).