inforMED
InjuryCDD

ACCESS VITAMIN B12 COBALAMIN

Received Aug 2, 2021 · Event occurred Jul 9, 2021

Report 9680746-2021-00050 · MDR key 12263693

Device

Generic name

Radioassay, Vitamin B12

Model number

33000

Catalog number

33000

Lot number

172011

Product problems

  • Adverse Event Without Identified Device or Use Problem
  • Adverse Event Without Identified Device or Use Problem

Patient

Not reported

  • Insufficient Information
  • Insufficient Information

Narrative

Additional Manufacturer Narrative

(B)(6). THE ACCESS B12 REAGENT WAS NOT RETURNED FOR EVALUATION. NO HARDWARE ERRORS, FLAGS OR OTHER ASSAY ISSUES WERE REPORTED IN CONJUNCTION WITH THIS EVENT. THERE IS INSUFFICIENT EVIDENCE TO SUGGEST A REAGENT OR HARDWARE MALFUNCTION OCCURRED IN THIS EVENT. IN CONCLUSION, THE CAUSE OF THIS EVENT CANNOT BE DETERMINED WITH THE AVAILABLE INFORMATION.

Description of Event or Problem

ON (B)(6) 2021 THE CUSTOMER REPORTED THAT ON (B)(6) 2021, ERRONEOUS LOW B12 RESULTS (ACCESS VITAMIN B12, PART NUMBER 33000 AND LOT NUMBER 172011) WERE GENERATED ON THE CUSTOMER'S DXI 800 (DXI 800 ACCESS IMMUNOASSAY ANALYZER W/ SPOT B, PART NUMBER A71456 AND SERIAL NUMBER (B)(4)) FOR AN UNKNOWN NUMBER OF PATIENTS. THE INITIAL LOW RESULTS WERE RELEASED FROM THE LABORATORY. THE CUSTOMER REPORTED A CHANGE TO PATIENT TREATMENT FOR SEVERAL PATIENTS; THE PATIENTS WERE PRESCRIBED VITAMIN B 12 SUPPLEMENTS (DOSAGES NOT PROVIDED). THE CUSTOMER REPORTED THE LOW RESULTS WERE DISCORDANT WITH PATIENTS¿ CLINICAL FILES. CUSTOMER ALSO REPORTED HIGHER B12 RESULTS WERE OBTAINED UPON REPEAT FOR SOME PATIENT SAMPLES (NO DATA PROVIDED). SYSTEM PERFORMANCE INDICATORS SUCH AS SYSTEM CHECK, CALIBRATIONS AND QUALITY CONTROL (QC) HAVE BEEN PASSING WITHIN SPECIFICATIONS AS REPORTED BY THE CUSTOMER. THERE WERE NO ISSUES WITH SAMPLE INTEGRITY REPORTED BY THE CUSTOMER. SAMPLE COLLECTION AND HANDLING INFORMATION SUCH AS SAMPLE TYPE, VOLUME DRAWN, TRANSPORTATION AND HANDLING, CENTRIFUGATION, STORAGE OR OTHER SAMPLE HANDLING INFORMATION WAS NOT PROVIDED BY THE CUSTOMER.