TINA-QUANT FERRITIN GEN.4
Received Jul 29, 2021 · Event occurred Jul 8, 2021
Report 1823260-2021-02205 · MDR key 12244520
Device
Generic name
Ferritin Immunological Test System
Manufacturer
Roche DiagnosticsModel number
NACatalog number
04885317190
Lot number
538712
Product problems
- High Test Results
- Non Reproducible Results
- High Test Results
- Non Reproducible Results
Patient
78 YR · Male
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
MEDWATCH FIELD D4 CATALOG NUMBER HAS BEEN UPDATED.
Additional Manufacturer Narrative
THE SAMPLE WAS REQUESTED FOR INVESTIGATION. THE INVESTIGATION IS ONGOING.
Description of Event or Problem
THE INITIAL REPORTER RECEIVED A QUESTIONABLE FERR4 TINA-QUANT FERRITIN GEN.4 RESULT FOR ONE PATIENT SAMPLE WITH THE COBAS 6000 C 501 MODULE SERIAL NUMBER (B)(4). THE INITIAL RESULT WAS 495 NG/ML. THE REPEATED RESULT RUN IN REDUCED VOLUME WAS 147 NG/ML. THE QUESTIONABLE RESULT WAS NOT REPORTED OUTSIDE OF THE LABORATORY. THE CUSTOMER IS QUESTIONING AN INTERFERENCE DUE TO PREVIOUS RESULTS SHOWING THE SAME PATTERN OF AN INITIALLY HIGH RESULT REPEATING LOWER.
Additional Manufacturer Narrative
THE SAMPLE WAS RECEIVED FOR INVESTIGATION AND TESTED USING A C501 AND AN E601. THE CUSTOMER¿S RESULTS COULD BE REPRODUCED. A DIFFERENCE BETWEEN THE FERRITIN CONCENTRATION VALUE GENERATED ON C501 IN NORMAL MODE AND THE FERRITIN VALUE GENERATED ON E601 COULD BE ALSO SEEN CLEARLY. THE RESULTS INDICATE AN INTERFERENT IN THE SAMPLE THAT INTERFERES WITH THE CLINICAL CHEMISTRY ASSAY FORMAT BUT HAS NO INFLUENCE ON THE ELECSYS ASSAY RESULT. THE PATIENT¿S MEDICATIONS WERE FURTHER INVESTIGATED FOR POTENTIAL INTERFERENCE. NONE OF THE TESTED ACTIVE COMPOUNDS OF THE MEDICATIONS SHOWED INTERFERENCE IN THE ASSAY. A FURTHER, MORE DETAILED CHARACTERIZATION OF THE POTENTIAL INTERFERING COMPOUND IS NOT POSSIBLE WITH THE REMAINING SAMPLE MATERIAL. THE INVESTIGATION FOUND THE ISSUE TO BE CONSISTENT WITH A SAMPLE-SPECIFIC UNKNOWN INTERFERENT. A PRODUCT PROBLEM COULD NOT BE FOUND.