EEG-1200A
Received Jul 24, 2021 · Event occurred Jun 24, 2021
Report 2080783-2021-01464 · MDR key 12222736
Device
Generic name
Neurofax Electroencephalograph
Manufacturer
Nihon Kohden CorporationModel number
EEG-1200ACatalog number
EEG-1200A
Lot number
NA
Product problems
- Overheating of Device
- Temperature Problem
- Overheating of Device
- Temperature Problem
Patient
Not reported
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE MONITOR TECHNICIAN REPORTED THAT THERE WAS A BURNING SMELL FROM THE SYSTEM. THEY DID SOME TROUBLESHOOTING AND FOUND THAT THE POWER SUPPLY INSIDE THE COMPUTER WAS CAUSING THE SMELL. THEY REPLACED THE PART WITH THE NIHON KOHDEN PROVIDED PART NUMBER FOR THE REPLACEMENT POWER SUPPLY. ISSUE WAS RESOLVED BY THE CUSTOMER. THIS WAS NOT IN PATIENT USE AND NO OTHER DEVICES BEING USED WITH THE SYSTEM.
Description of Event or Problem
THE MONITOR TECHNICIAN REPORTED THAT THERE WAS A BURNING SMELL FROM THE SYSTEM. THEY DID SOME TROUBLESHOOTING AND FOUND THAT THE POWER SUPPLY INSIDE THE COMPUTER WAS CAUSING THE SMELL. THEY REPLACED THE PART WITH THE NIHON KOHDEN PROVIDED PART NUMBER FOR THE REPLACEMENT POWER SUPPLY. ISSUE WAS RESOLVED BY THE CUSTOMER. THIS WAS NOT IN PATIENT USE.