Received Jul 15, 2021 · Event occurred Jun 22, 2021

Report 1717344-2021-00984 · MDR key 12178873

Device

Generic name

Counter, Sponge, Surgical

Model number

01-0043

Catalog number

01-0043

Product problems

  • False Positive Result
  • False Positive Result

Patient

Not reported

  • Unintended Radiation Exposure
  • Unintended Radiation Exposure

Narrative

Description of Event or Problem

ACCORDING TO THE REPORTER, DURING A LEFT NECK DISSECTION, UNTRAINED NURSE WAS USING THE SITUATE ROOM SCANNER TO SCAN THE PATIENT. CUSTOMER STATED THAT THE WAND WAS INDICATING A DETECTION WHEN SCANNING OVER THE UPPER RIGHT CHEST BUT NOWHERE ELSE. THIS RESULT WAS NOT REPRODUCIBLE ON ANY OTHER AREA OF THE BODY. IN CONSULTING WITH THE SITUATE CLINICAL SUPPORT SPECIALIST AND CUSTOMER, IT WAS DETERMINED THAT THE POSITIVE SCAN WAS MOST LIKELY OPERATOR ERROR WITH AN RF DETECTABLE SPONGE TOO CLOSE TO THE SURGICAL FIELD. X-RAY WAS DONE TO CONFIRM IF THERE WAS NO RETAINED SPONGE. NO OTHER UNIT WAS USED.

Additional Manufacturer Narrative

H3 EVALUATION SUMMARY: MEDTRONIC CONDUCTED AN INVESTIGATION BASED UPON ALL INFORMATION RECEIVED. THE DEVICE WAS AVAILABLE FOR EVALUATION. THE EVALUATION FOUND THAT LEFT ACCESSORY PORT WAS DIFFICULT TO REMOVE DURING TESTING AND THE RIGHT PORT FOR THE MAT WAS LOOSED. IT WAS REPORTED THAT THE DEVICE GAVE A FALSE POSITIVE DETECTION. THE REPORTED ISSUE COULD NOT BE CONFIRMED. THE MOST LIKELY CAUSE COULD NOT BE ESTABLISHED FROM THE INFORMATION AVAILABLE. THE EVALUATION DETECTED AN UNREPORTED CONDITION: THE UNIT HAD A CONNECTOR FAILURE. THE MOST LIKELY CAUSE FOR THE ADDITIONAL CONDITION IS TRACED TO A COMPONENT FAILURE. THE MANUFACTURING RECORDS FOR EACH DEVICE ARE THOROUGHLY REVIEWED PRIOR TO RELEASE TO ENSURE THAT IT MEETS ALL MEDTRONIC QUALITY SPECIFICATIONS. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem

ACCORDING TO THE REPORTER, DURING A PROCEDURE, UNTRAINED NURSE WAS USING THE SITUATE ROOM SCANNER TO SCAN THE PATIENT. CUSTOMER STATES THAT THE WAND WAS INDICATING A DETECTION WHEN SCANNING OVER THE UPPER RIGHT CHEST, BUT NOWHERE ELSE. THIS RESULT WAS NOT REPRODUCIBLE ON ANY OTHER AREA OF THE BODY. IN CONSULTING WITH THE SITUATE CLINICAL SUPPORT SPECIALIST AND CUSTOMER, IT WAS DETERMINED THAT THE POSITIVE SCAN WAS MOST LIKELY OPERATOR ERROR WITH AN RF DETECTABLE SPONGE TOO CLOSE TO THE SURGICAL FIELD. X-RAY WAS DONE TO CONFIRM IF THERE WAS NO RETAINED SPONGE.