FIBERSCOPE, CYSTO-URETHRO, REVERSE
Received Jul 2, 2021
Report 1221826-2021-00054 · MDR key 12108250
Device
Product problems
- Material Puncture/Hole
- Material Perforation
- Material Integrity Problem
- Material Puncture/Hole
- Material Perforation
- Material Integrity Problem
- Adverse Event Without Identified Device or Use Problem
Patient
Unknown
- Unspecified Infection
- Unspecified Infection
Narrative
Additional Manufacturer Narrative
THE ITEM HAS RETURNED. SEVERAL UNSUCCESSFUL ATTEMPTS HAVE BEEN MADE TO GATHER FURTHER INFORMATION FROM THE CLINIC. ONCE INFORMATION IS RECEIVED, IT WILL DETERMINE IF A MICROBIOLOGICAL ASSESSMENT NEEDS TO BE PERFORMED. A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF EVALUATION.
Description of Event or Problem
ALLEGEDLY, THE FLEXIBLE CYSTOSCOPE WAS USED IN CASES AND THE CLINIC REPORTS PATIENTS CONTRACTED AN INFECTION.
Additional Manufacturer Narrative
TESTING PERFORMED BY GLOBAL RSB SHOWED NO VIABLE SIGNS OF MICROORGANISMS NOR BIOBURDEN IN EXTRACT SAMPLES FROM THE WORKING CHANNEL FOR ITEM 11272CK2 / SN (B)(6): 1. INITIAL EXTRACTION OF THE WORKING CHANNEL UPON DEVICE WAS RECEIVED. 2. REPROCESSING THE ENDOSCOPE MANUALLY ACCORDING TO OUR IFU (WITHOUT STERILIZATION). 3. 2ND EXTRACTION OF WORKING CHANNEL. THE RESULTS FROM THE LABORATORY SHOW NO VIABLE MICROORGANISM IN BOTH EXTRACTS: NO CONTAMINATION / BIOBURDEN DETECTABLE IN THE CHANNEL. ITEM WAS THEN RETURNED TO MANUFACTURE ON 10/13/2021 FURTHER EVALUATION FOUND BROKEN IMAGE FIBERS, WHICH IS THE REASON FOR THE LOW LIGHT OUTPUT; 30.4MW, DEBRIS ON VERTEBRAE, EXTERNAL PUNCTURE IN THE ANGLE COVER AT 10MM FROM DISTAL END, LEAKING AND DEFLECTION FAILED; 210° UP/115° DOWN.