inforMED
MalfunctionFWC

NICVIEW 2.0 CAMERA

Received Jun 28, 2021 · Event occurred Jun 11, 2021

Report 9612330-2021-00003 · MDR key 12077067

Device

Generic name

Nicview 2.0 Camera

Model number

NVCAM

Catalog number

NVCAM

Product problems

  • Break
  • Connection Problem
  • Break
  • Connection Problem

Patient

Not reported

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

FOLLOW UP REPORT 001 REF NATUS COMPLAINT# (B)(4). CORRECTION TO SECTION B1., G2., H1. SECTION D4., NO UNIQUE IDENTIFIER (UDI) DOCUMENTED AS NOT APPLICABLE. THE AFFECTED UNIT WAS RECEIVED INTO MIDDLETON. EXAMINATION AND TESTING WAS NOT PERFORMED BECAUSE FACTORY SERVICE DOES NOT SERVICE NICVIEW 2 PRODUCTS. CAPA005201 IS RELATED TO THIS COMPLAINT. CORECTIVE ACTION PLAN: NICVIEW 2.0 - IMPROVEMENT TO THE MICROUSB (POWER CABLE END) AT THE BASE OF THE NICVIEW 2 ARM. NICVIEW 2.0 - IMPROVEMENT TO THE MICROUSB (POWER CABLE) AT CAMERA END CREATED TO DESIGN PROTECTION FOR THE MICROUSBS. RISK ASSESSMENT WAS CONDUCTED AND RELEASED ON DOC-055534 REV 01 NICVIEW 2 MICRO-USB DAMAGE POST MARKET RISK ASSESSMENT. PREVENTIVE ACTION PLAN: INSTALL BRACKET, PN 036431, ON THE ARM USING SCREWS INSTEAD OF ADHESIVE AT THE MANUFACTURER. INVESTIGATION RESULT CODE: NEURO SBU/CONNECTOR DAMAGED THIS COMPLAINT WAS VERIFIED, CAPA005201 ALREADY OPEN. BASED ON EVALUATION OF THIS COMPLAINT THERE IS NO SAFETY RISK OF HARM, INDIVIDUAL COMPLAINT RELATED TO ISSUE STATED. COMPLAINT WILL BE INCLUDED IN TRENDING DATA FOR FURTHER REVIEW AND INVESTIGATION IF NEEDED. PER QMS-004442 COMPLAINT HISTORIES ARE REVIEWED ROUTINELY PER QUALITY SYSTEM REQUIREMENTS AND ANY COMPLAINT TRENDS ARE ASSESSED AND DOCUMENTED AS PART OF THESE REVIEWS. PER DOC-023354 LINE 2.2 SEVERITY 11- ACCEPTABLE RESIDUAL RISK LEVEL. A RISK REVIEW IS NOT REQUIRED AS THIS COMPLAINT DOES NOT DESCRIBE A NEW FAILURE MODE OR NEW HARM AND THE EXISTING HAZARD SEVERITY AND/OR PROBABILITY OF OCCURRENCE HAS NOT CHANGED. A SEARCH FOR SERVICE DATA THAT MAY HAVE BEEN RELEVANT TO THIS ISSUE WAS CONDUCTED. NO FURTHER INFORMATION RELATED TO THIS ISSUE WAS FOUND.

Description of Event or Problem

DAMAGED CAMERA/PWRSUPPLY CONNECTION. (NICVIEW 2.0 CAMERA) SPARKS WHEN TRYING TO PLUG BACK IN. NO PATIENT INVOLEMENT.

Additional Manufacturer Narrative

INITIAL REPORT REF NATUS COMPLAINT# (B)(4). CAM #4, MAC: B8:27:EB:90:E8:4C, DAMAGED CAMERA/PWRSUPPLY CONNECTION SPARKS WHEN TRYING TO PLUG BACK IN. NO PATIENT INJURY. DAMAGED NVCAM RETURNING BACK TO MIDDLETON FOR EVALUATION.

Description of Event or Problem

DAMAGED CAMERA/PWRSUPPLY CONNECTION. (NICVIEW 2.0 CAMERA) SPARKS WHEN TRYING TO PLUG BACK IN. NO PATIENT INVOLVEMENT.