COMPAX 40E
Received Oct 20, 2008 · Event occurred Aug 16, 2008
Report 2126677-2008-00085 · MDR key 1204588
Device
Manufacturer
Ge Medical Systems, LlcProduct problems
- Unstable
- Device Operates Differently Than Expected
- Unintended Movement
Patient
Not reported
- No Known Impact Or Consequence To Patient
Narrative
Additional Manufacturer Narrative
THE GE FIELD ENGINEER (FE) FOUND A BLOWN 2-AMP FUSE. WHEN THE FE REPLACED THE FUSE, IT WAS VERIFIED THAT THE LOCKS WERE PERFORMING ACCORDING TO SPECS.
Description of Event or Problem
IT WAS REPORTED THAT THE TABLE LOCKS DID NOT ACTUATE, CAUSING THE TABLETOP TO UNEXPECTEDLY MOVE IN BOTH THE LONGITUDINAL AND LATERAL DIRECTIONS WITHOUT RESISTANCE (FREE FLOAT). THE PROBLEM WAS DISCOVERED AS THE OPERATOR WAS POSITIONING THE PT. THERE WAS NO REPORT OF A FALL OR INJURY. THIS SITUATION COULD CONTRIBUTE TO AN INJURY IF A PT OR OPERATOR WERE UNAWARE OF THIS CONDITION WHILE LOADING OR UNLOADING A PT. THE ENSUING INSTABILITY COULD LEAD TO A FALL.