MalfunctionIZZ

COMPAX 40E

Received Oct 16, 2008 · Event occurred Sep 17, 2008

Report 2126677-2008-00084 · MDR key 1203586

Device

Model number

45435245

Catalog number

NA

Lot number

NA

Product problems

  • Unstable
  • Device Operates Differently Than Expected
  • Unintended Movement

Patient

NA

  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THE TABLE LOCKS DID NOT ACTUATE, THAT CAUSED THE TABLETOP TO UNEXPECTEDLY MOVE IN BOTH LONGITUDINAL AND LATERAL DIRECTIONS WITHOUT RESISTANCE (FREE FLOAT). THERE WAS NO INJURY REPORTED. THIS SITUATION COULD CONTRIBUTE TO AN INJURY IF A PT OR OPERATOR WERE UNAWARE OF THIS CONDITION WHILE LOADING OR UNLOADING A PT. THE ENSUING INSTABILITY COULD LEAD TO A FALL.

Additional Manufacturer Narrative

THE GE FIELD ENGINEER (FE) FOUND A MALFUNCTION IN THE MAIN POWER SUPPLY BOARD. THE FE REPLACED THE BOARD AND TESTED THE SYSTEM. THE TABLELOCKS WERE VERIFIED TO BE PERFORMING WITHIN SPECIFICATIONS.