inforMED
MalfunctionBXJ

NEOTECH PRODUCTS

Received Jun 10, 2021

Report 2025917-2021-00005 · MDR key 11974562

Device

Generic name

Neoseal

Model number

N422

Catalog number

N422

Lot number

2020-0813

Product problems

  • Loss of or Failure to Bond
  • Loss of or Failure to Bond

Patient

Not reported

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

DEVICE WAS NOT RETURNED AND NO EVALUATION WAS POSSIBLE ON THE ACTUAL DEVICE. HOWEVER, SAMPLES OF FINISHED GOODS WERE PULLED OUT FROM RETAINED SAMPLES OF BOTH REPORTED AND CURRENT LOT NUMBERS FOR RE-INSPECTION AND NO ABNORMALITY WAS FOUND. ALSO, THE PRODUCT'S DEVICE HISTORY RECORDS (DHR) OF REPORTED LOT AND CURRENT INVENTORY WERE REVIEWED AND NO NONCONFORMITY WAS FOUND. PER OUR INVESTIGATION DEVICE WAS USED WITH RAM CANNULA AND ON OFF LABEL USE FOR CPAP. THIS COULD BE CONTRIBUTED TO THESE ISOLATED INCIDENTS AS WE JUST HAD ONE SIMILAR INCIDENT FROM THE SAME SOURCE IN THE PAST FIVE YEARS FOR MILLIONS OF THE PARTS THAT WERE USED ALL AROUND THE WORLD. PER COMPLAINTS RATIO, THESE INCIDENTS WERE ISOLATED EVENTS. THIS COMPLAINT WILL BE MONITORED FOR TRENDING PURPOSES AND IS ADDED TO THE RELATED LOGS AND CHARTS.

Description of Event or Problem

PER CUSTOMER, NEOSEAL FOUND IN BABY'S MOUTH BUT WAS REMOVED IN TIME.

Remedial action

  • Inspection