inforMED
MalfunctionCAP

AIRWAY PRESSURE MONITOR

Received Sep 12, 1997 · Event occurred Aug 13, 1997

Report 2518422-1997-00019 · MDR key 119185

Device

Generic name

Airway Pressure Monitor

Manufacturer

Respironics, Inc.

Model number

332259

Catalog number

332259

Lot number

NA

Product problems

Patient

NA

  • No Consequences Or Impact To Patient

Narrative

Description of Event or Problem

DURING EVALUATION AND REPAIR OF RETURNED DEVICE, IT WAS NOTED THAT THE LOW ALARM WOULD NOT TRIGGER UNTIL THE GAUGE WAS IN THE NEGATIVE RANGE, AT ABOUT -10. THE CUSTOMER WAS THEN CONTACTED FOR FURTHER INFO. THE CUSTOMER INDICATED THAT THE PROBLEM WAS DISCOVERED DURING THEIR INTERNAL TESTING. THE DEIVICE WAS NOT BEING USED BY A PT AT THE TIME.

Remedial action

  • Repair