STATLOCK SLIDING POST CRESCENT PAD SKIN PREP
Received May 20, 2021 · Event occurred Jan 14, 2021
Report 3006260740-2021-01900 · MDR key 11856207
Device
Generic name
Stand, Infusion
Manufacturer
C.r. Bard, Inc. (basd) -3006260740Catalog number
PIC0220CE
Lot number
JUEP2260
Product problems
- Detachment of Device or Device Component
- Detachment of Device or Device Component
Patient
Not reported
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE FOLLOWING WERE REVIEWED AS PART OF THIS INVESTIGATION: PATIENT SEVERITY, COMPLAINT AND LOT HISTORY REVIEW, APPLICABLE PREVIOUS INVESTIGATION(S), SAMPLE (IF AVAILABLE), APPLICABLE MANUFACTURE RECORDS, AND APPLICABLE FMEA DOCUMENTS. BASED ON A REVIEW OF THIS INFORMATION, THE FOLLOWING WAS CONCLUDED: THE COMPLAINT OF A DETACHED RETAINER IS CONFIRMED AND WAS DETERMINED TO BE MANUFACTURING RELATED. TWO PICC PLUS STATLOCK KITS WERE RETURNED FOR EVALUATION. AN INITIAL VISUAL OBSERVATION SHOWED ONE OF THE KITS WAS RETURNED OPEN AND ONE KIT WAS RETURNED SEALED. THE RETAINER OF THE STATLOCK DEVICE RETURNED IN THE OPEN KIT WAS FOUND TO BE COMPLETELY DETACHED FROM THE PAD, AND THE UNDERSIDE OF THIS DETACHED RETAINER WAS OBSERVED TO BE TACKY. A MICROSCOPIC OBSERVATION REVEALED GOOD ADHESIVE COVERAGE ON THE UNDERSIDE OF BOTH RETAINERS; HOWEVER, STRINGS OF ADHESIVE WERE FOUND ON THE UNDERSIDE OF THE RETAINER OF THE SEALED SAMPLE WHEN AN ATTEMPT WAS MADE TO REMOVE IT FROM THE PAD. THIS SUGGESTS THAT THE ADHESIVE BETWEEN THE RETAINER AND PAD MAY NOT HAVE BEEN FULLY CURED. THE INVESTIGATION HAS BEEN FORWARDED TO THE MANUFACTURING FACILITY FOR FURTHER EVALUATION, AND BD IS WORKING CLOSELY WITH THE MANUFACTURING FACILITY TO PREVENT RECURRENCE OF THE REPORTED EVENT. A LOT HISTORY REVIEW (LHR) OF JUEP2260 SHOWED ONE OTHER SIMILAR PRODUCT COMPLAINT(S) FROM THIS LOT NUMBER. THE COMPLAINTS FOR THIS LOT NUMBER (JUEP2260) HAVE BEEN REPORTED FROM MULTIPLE FACILITIES IN (B)(6).
Description of Event or Problem
IT WAS REPORTED THE PLASTIC CLAMPS ON THE STATLOCK DEVICE FALLS OFF THE BAND AID SO THE PICC CATHETER IS AT RISK FOR SLIDING OUT OFF THE PATIENT BECAUSE IT IS NOT FIXED. SO PLASTIC CLAMPS ARE NOT FIXED CORRECT ON THE BAND AID. NO OTHER INFORMATION WAS PROVIDED. 04/21/2021- TWO DEVICES WERE RETURNED. THIS REPORT ADDRESSES THE SECOND DEVICE.