ENDO DISSECT
Received May 6, 2021 · Event occurred Apr 13, 2021
Report 2647580-2021-01586 · MDR key 11783604
Device
Generic name
Motor, Surgical Instrument, Pneumatic Powered
Manufacturer
Us Surgical Puerto RicoModel number
176645Catalog number
176645
Lot number
P8A0575X
Product problems
- Break
- Detachment of Device or Device Component
- Break
- Detachment of Device or Device Component
Patient
Not reported
- Unintended Radiation Exposure
- Unintended Radiation Exposure
Narrative
Additional Manufacturer Narrative
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
Description of Event or Problem
ACCORDING TO THE REPORTER, WHILE MOBILIZING TISSUE DURING A LAPAROSCOPIC CHOLECYSTECTOMY, ONE SIDE OF THE JAWS BROKE OFF AND FELL INTO THE ABDOMEN. IT TOOK AN X-RAY AND 45 MINUTES TO LOCATE THE BROKEN OFF PART OF THE DEVICE AND SURGEON RETRIEVED IT. ANOTHER DEVICE WAS OPENED AND USED TO COMPLETE THE CASE.
Additional Manufacturer Narrative
EVALUATION SUMMARY: MEDTRONIC CONDUCTED AN INVESTIGATION BASED UPON ALL INFORMATION RECEIVED. THE DEVICE WAS AVAILABLE FOR EVALUATION. VISUAL INSPECTION NOTED THE JAW OF THE DEVICE WAS BROKEN AT THE FULCRUM. IT WAS REPORTED THAT THE JAWS OF THE DEVICE WERE BROKEN, A COMPONENT DISENGAGED FROM THE DEVICE INTO THE SURGICAL CAVITY, AND OR TIME WAS EXTENDED BY >30 MINUTES RESULTING FROM PRODUCT FAILURE. THE REPORTED ISSUES WERE CONFIRMED. THE PRODUCT ANALYSIS NOTED EVIDENCE THAT THE DEVICE WAS NOT USED AS INTENDED. BROKEN JAW MAY OCCUR WHEN THE DEVICE IS EXPOSED TO A SIDE FORCE (LEVERAGE) THAT CONSEQUENTLY BREAKS ONE SIDE OF THE JAWS. ANOTHER POSSIBILITY IS WHEN THE DEVICE IS ACTIVATED WITH TOO MUCH FORCE TO THE JAWS AND IS USED IN A TWISTING MOTION RESULTING IN BREAKAGE. THE MANUFACTURING RECORDS FOR EACH DEVICE ARE THOROUGHLY REVIEWED PRIOR TO RELEASE TO ENSURE THAT IT MEETS ALL MEDTRONIC QUALITY SPECIFICATIONS. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.